Instructions

Clear and sanitize procedures
Memory devices
The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options. Detailed procedures to
clear or sanitize these devices, if any, are given following each table.
Terminology
The following terms are used in the tables in this section:
User data – Describes the type of information stored in the device. Refers to waveforms or other measurement data representing
signals connected to the instrument by users.
User settings – Describes the type of information stored in the device. Refers to instrument settings that can be changed by the user.
Both – Describes the type of information stored in the device. It means that both user data and user settings are stored in the device.
None – Describes the type of information stored in the device. It means that neither user data nor user settings are stored in the
device.
Directly – Describes how data is modified. It means that the user can modify the data.
Indirectly – Describes how data is modified. It means that the instrument system resources modify the data and that the user cannot
modify the data.
Volatile memory devices
These are the memory capacities at the time of publishing this document, but are subject to change.
Type and
size
Function Type of
user info
stored
Backed up
by battery
Method of
modificatio
n
Data input
method
Location User
accessible
To clear To sanitize
SDRAM 2
GB (All
models)
System
memory
Both No Indirectly Written by
processor
system and
application
software
operations
Main board No Unplug the
instrument
and remove
all battery
packs for at
least 30
seconds
Unplug the
instrument
and remove
all battery
packs for at
least 30
seconds
CMOS RAM
13 bytes
Holds clock
and boot
configuratio
n data
None Yes Indirectly Boot
operations
Main board No Cannot be
cleared
Remove the
Main board
FPGA <20
MB
Acquisition
system
None No None Written by
processor
Main board No Unplug the
instrument
and remove
all battery
packs for at
least 30
seconds
Unplug the
instrument
and remove
all battery
packs for at
least 30
seconds
Clear and sanitize procedures
2 Series Mixed Signal Oscilloscope MSO22, MSO24 Declassification and Security Instructions 5