Datasheet

Vertical system digital channels
Input channels 16 digital (D15 to D0)
Thresholds Threshold per set of 8 channels
Threshold selections TTL, CMOS, ECL, PECL, User-defined
User-defined threshold range ± 20 V
Threshold accuracy ±[100 mV + 3% of threshold setting]
Maximum input voltage ± 40 V
Input dynamic range 80 V
pk-pk
(threshold setting dependent)
Minimum voltage swing 500 mV
p-p
Input resistance 101 kΩ
Probe loading 8 pF
Vertical resolution 1 bit
Horizontal system analog channels
Maximum duration of time
captured at highest sample rate
(all channels)
1 ms
Time base range 200 MHz models: 2ns to 100s
70, 100 MHz models: 4ns to 100s
Time-base delay time range -10 divisions to 5000 s
Channel-to-channel deskew range ±100 ns
Time base accuracy ±25 ppm over any ≥1 ms interval
Horizontal system digital channels
Maximum sample rate (main)
When using any of channels D7 - D0
When using any of channels D8 - D15
1 GS/s (1 ns resolution) 500 MS/s (2 ns resolution)
Maximum record length (main, all
channels)
1 M points
Minimum detectible pulse width 5 ns
Channel-to-channel skew 2 ns typical
MSO/DPO2000B Series Mixed Signal Oscilloscope
www.tek.com 9