Datasheet

Mixed Domain Oscilloscopes MDO4000 Series
Display Characteristics
Characteristic
Description
Display Type
10.4 in. (264 mm) liquid-crystal TFT color display
Display Resolution
1,024 horizontal × 768 vertical pixels (XGA)
Waveform Styles Vectors, Dots, Variable Persistence, Innite
Persistence
Graticules Full, Grid, Solid, Cross Hair, Frame, IRE, and mV
Format
YT and simultaneous XY/YT
Waveform Capture Rate >50,000 wfm /s maximum
Input/Output Ports
Port Description
USB 2.0 High-speed Host
Port
Supports USB mass storage devices, printers and
keyboard. Two ports on front and two ports on rear of
instrument
USB 2.0 Device Port Rear-panel connector allows for communication/control
of oscilloscope through USBTMC or GPIB
(with a TEK-USB-488), and direct printing to all
PictBridge-compatible printers
LAN Port
RJ-45 connector, supports 10/100/1000 Mb/s
XGA Video Port DB-15 female connector, connect to show the
oscilloscope display on an external monitor or projector
Probe Compensator
Output
Front-
panel pins
Amplit
ude: 2.5 V
Freque
ncy: 1 kHz
Auxiliary Out Rear-panel BNC connector
V
OUT
(Hi): 2.5 V open circuit, 1.0 V 50 to ground
V
OUT
(Lo): 0.7Vintoaloadof4mA;0.25 V 50
to ground
Outputcanbecongured to provide a pulse out signal
when the oscilloscope triggers, the internal oscilloscope
reference clock out, or an event out for limit/mask
testing
Ex
ternal Reference In
Time-base system can phase lock to an external
10 MHz reference (10 MHz ±1%)
Kensington Lock Rear-panel security slot connects to standard
Kensington lock
VESA Mount Standard (MIS-D 100) 100 mm VESA mounting points
on rear of instrument
LAN eXtensions for Instrumentation (LXI)
Characteristic
Description
Class LXI Class C
Version V1.3
Power Source
Characteristic
Description
Power Source Voltage 100 t o 240 V ±10%
Power Source Frequency 45 to 66 Hz (85 to 264 V)
360 to 440 Hz (100 to 132 V)
Power Consumption
225 W maximum
Physical Characteristics
Dimensions
mm
in.
Height 229 9.0
Width 439 17.3
Depth 147 5.8
Weight kg lb.
Net
5
11
Shipping
10.7 23.6
Rackmount Conguration
5U
Cooling Clearance 2 in. (51 mm) required on left side and rear of
instrument
Environmental
Characteristic
Description
Temperature
Operating 0 ºC to +50 º C
Nonoperating
–20 ºC to +60 ºC
Humidity
Operating High: 40 ºC to 50 ºC, 10% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 10% to 90% Relative Humidity
Nonoperating
High: 40 ºC to 60 ºC, 5% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 5% to 90% Relative Humidity
Altitude
Operating 9,843 ft. (3,000 m)
Nonoperating
30,000 ft. (9,144 m)
Regulatory
Electromagnetic
Compatibility
EC Council Directive 2004/108/EC
Safety UL61010-1, Second Edition; CSA61010-1 Second
Edition, EN61010-1:2001; IEC 61010-1:2001
Ordering Information
MDO4000 Family
Model Description
MD
O4054-3
Mi
xed Domain Oscilloscope with (4) 500 MHz analog
ch
annels, (16) digital channels, and (1) 3 GHz RF input
MDO4054-6 Mixed Domain Oscilloscope with (4) 500 MHz analog
channels, (16) digital channels, and (1) 6 GHz RF input
MDO4104-3 Mixed Domain Oscilloscope with (4) 1 GHz analog
channels, (16) digital channels, and (1) 3 GHz RF input
MDO4104-6 Mixed Domain Oscilloscope with (4) 1 GHz analog
channels, (16) digital channels, and (1) 6 GHz RF input
All Models Include: One passive voltage probe per analog channel (TPP0500
500 MHz, 10X, 3.9 pF for 500 MHz models; TPP1000 1 GHz, 10X, 3.9 pF for
1 GHz models), P6616 16-channel Logic Probe, Logic Probe Accessory Kit
(020-2662-xx), Front Cover (200-5130-xx), N-to-BNC Adapter (103-0045-00), User
Manual (071-2918-xx), Documentation CD (063-4367-xx), O penChoice
®
Desktop
Software, NI LabVIEW SignalExpress™ Tektronix Edition Software, Calibration
Certicates document measurement traceability to National Metrology Institute(s)
the Quality System this product is manufactured in is ISO9001 registered, power
cord, accessory bag (016-2030-xx) and a three-year warranty. Please specify power
plug and manual language version when ordering.
www.tektronix.com 17