Datasheet
Mixed Domain Oscilloscopes — MDO4000 Series
Characteristic MDO4054-3 MDO4104-3 MDO4054-6 MDO4104-6
Frequency Domain Trace
Types
Normal,Average,MaxHold,MinHold
Time Domain Trace Types Amplitude vs. Time, Frequency vs. Time, Phase vs. Time
Detection Methods
+Peak, –Peak, Average, Sample
Automatic Markers
1-11 peaks identified based on user-adjustable threshold and excursion values
Manual Markers
Two manual markers indicating frequency, amplitude, noise density, and phase noise
Marker Readouts Absolute or Delta
RF Acquisition Length
Span
Maximum R F Acquisition Time
>2 GHz
2.5 ms
>1 GHz - 2 GHz
5ms
>800 MHz - 1 GHz
10 ms
>500 MHz - 800 MHz 12.5 ms
>400 MHz - 500 MHz 20 ms
>250 MHz - 400 MHz 25 ms
>200 MHz - 250 MHz 40 ms
>160 MHz - 200 MHz 50 ms
>125 MHz - 160 MHz 62.5 ms
<125 MHz 79 ms
FFT Windows
FFT Window Factor
Kaiser 2.23
Rectangular 0.89
Hamming 1.30
Hanning 1.44
Blackman-Harris 1.90
Flat-Top 3.77
www.tektronix.com 11