Datasheet

Datasheet
www.tektronix.com36
e*Scope
®
Web-based interface Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the
IP address or network name of the oscilloscope and a web page will be served to the browser. Transfer and save
settings, waveforms, measurements, and screen images or make live control changes to settings on the oscil-
loscope directly from the web browser.
LXI Web interface Connect to the oscilloscope through a standard Web browser by simply entering the oscilloscope’s IP address
or network name in the address bar of the browser. The Web interface enables viewing of instrument status and
configuration, status and modification of network settings, and instrument control through the e*Scope Web-
based remote control. All Web interaction conforms to LXI Core specification, version 1.4.
Power source
Power source voltage 100 to 240 V ±10%
Power source frequency 50 to 60 Hz ±10% at 100 to 240 V ±10%
400 Hz ±10% at 115 V ±13%
Power consumption 250 W maximum
Physical characteristics
Dimensions
mm in.
Height 229 9
Width 439 17.3
Depth 147 5.8
Weight
Instruments without option SA3 or SA6
kg lb
Net 5.5 12.2
Shipping 11.2 24.8
Instruments with option SA3 or SA6
kg lb
Net 5.1 11.2
Shipping 10.8 23.8
Rackmount conguration 5U
Cooling clearance 2 in. (51 mm) required on left side and rear of instrument
EMC environment and safety
Temperature
Operating 0 ºC to +50 ºC (+32 ºF to 122 ºF)
Nonoperating -30 ºC to +70 ºC (-4 ºF to 140 ºF)
Humidity
Operating High: 40 ºC to 50 ºC, 10% to 60% relative humidity, Low: 0 ºC to 40 ºC, 10% to 90% relative humidity
Nonoperating High: 40 ºC to 60 ºC, 5% to 55% relative humidity, Low: 0 ºC to 40 ºC, 5% to 90% relative humidity
Altitude
Operating 3,000 meters (9, 843 feet)
Nonoperating 12,000 meters (39, 370 feet)
Regulatory CE marked for the European Union and UL approved for the US and Canada