Datasheet
Datasheet
www.tektronix.com34
Ethernet
3
10BASE-T and 100BASE-TX: Trigger on Start Frame Delimiter, MAC Addresses, MAC Q-Tag Control Information,
MAC Length/
Type, IP Header, TCP Header, TCP/IPv4/MAC Client Data, End of Packet, and FCS (CRC) Error.
100BASE-TX: Idle.
MAC Addresses – Trigger on Source and Destination 48-bit address values. MAC Q-Tag Control Information –
Trigger on Q-Tag 32-bit value.
MAC Length/Type – Trigger on ≤, <, =, >, ≥, ≠ a particular 16-bit value, or inside or outside of a range. IP Header
– Trigger on IP Protocol 8-bit value, Source Address, Destination Address.
TCP Header – Trigger on Source Port, Destination Port, Sequence Number, and Ack Number.
TCP/IPv4/MAC Client Data – Trigger on ≤, <, =, >, ≥, ≠ a particular data value, or inside or outside of a range.
Selectable number of bytes to trigger on from 1-16. Byte offset options of Don’t Care, 0-1499.
CAN Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data,
Identifier and Data, End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s. Data can be
further specified to trigger on ≤, <, =, >, ≥, or ≠a specific data value. User-adjustable sample point is set to 50%
by default.
LIN Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, Errors such as Sync, Parity, or
Checksum Errors up to 100 kb/s (by LIN definition, 20 kb/s).
FlexRay Trigger on Start of Frame, Type of Frame (Normal, Payload, Null, Sync, Startup), Identifier, Cycle Count, Complete
Header Field, Data, Identifier and Data, End of Frame or Errors such as Header CRC, Trailer CRC, Null Frame,
Sync Frame, or Startup Frame Errors up to 100 Mb/s.
MIL-STD-1553 Trigger on Sync, Word Type
3
(Command, Status, Data), Command Word (set RT Address, T/R, Sub-address/
Mode, Data WordCount/Mode Code, and Parity individually), Status Word (set RT Address, Message Error,
Instrumentation, Service Request Bit, Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus
Control Acceptance (DBCA), Terminal Flag, and Parity individually), Data Word (user-specified 16-bit data value),
Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time (minimum time selectable from 2 µs to 100 µs;
maximum time selectable from 2 µs to 100 µs; trigger on < minimum, > maximum, inside range, outside range).
RT Address can be further specified to trigger on =, ≠, <, >, ≤, ≥ a particular value, or inside or outside of a range.
I
2
S/LJ/RJ/TDM Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific
data value, or inside or outside of a range. Maximum data rate for I
2
S/LJ/RJ is 12.5 Mb/s. Maximum data rate for
TDM is 25 Mb/s.
6. Digital Voltmeter and Frequency Counter
Source Channel 1, Channel 2, Channel 3, Channel 4
Measurement types AC RMS, DC, AC+DC RMS (reads out in volts or amps); Frequency
Voltage accuracy ±(1.5% |reading – offset – position|) + (0.5% |(offset – position)|) + (0.1 * Volts/div))
Resolution ACV, DCV: 4 digits
Frequency: 5 digits
Frequency accuracy ±(10 µHz/Hz + 1 count)
Measuring rate 100 times/second; measurements updated on the display 4 times/second
Vertical settings autorange Automatic adjustment of vertical settings to maximize measurement dynamic range; available for any non-
trigger source
Graphical measurement Graphical indication of minimum, maximum, current value, and five second rolling range
3 ≥350 MHz bandwidth models are recommended for 100BASE-TX