Datasheet
www.tektronix.com 31
Mixed Domain Oscilloscopes
Square/Pulse
Frequency range 0.1 Hz to 25 MHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Duty cycle 10% to 90% or 10 ns minimum pulse, whichever is larger cycle
Duty cycle resolution 0.1%
Pulse width minimum (typical) 10 ns
Rise/fall time (typical) 5 ns (10% –90%)
Pulse width resolution 100 ps
Overshoot (typical) < 2% for signal steps greater than 100 mV
Asymmetry ±1% ±5 ns, at 50% duty cycle
Jitter (TIE RMS) (typical) < 500 ps
Ramp/Triangle
Frequency range 0.1 Hz to 500 kHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Variable symmetry 0% to 100%
Symmetry resolution 0.1%
DC
Level range (typical) ±2.5 V into Hi-Z; ±1.25 V into 50 Ω
Noise
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z;10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Amplitude resolution 0% to 100% in 1% increments
Sin(×)/× (Sinc)
Frequency range (typical) 0.1 Hz to 2 MHz
Amplitude range 20 mV
p-p
to 3.0 V
p-p
into Hi-Z; 10 mV
p-p
to 1.5 V
p-p
into 50 Ω
Gaussian
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.5 V
p-p
into Hi-Z; 10 mV
p-p
to 1.25 V
p-p
into 50 Ω
Lorentz
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.4 V
p-p
into Hi-Z; 10 mV
p-p
to 1.2 V
p-p
into 50 Ω
Exponential Rise/Decay
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.5 V
p-p
into Hi-Z; 10 mV
p-p
to 1.25 V
p-p
into 50 Ω
Haversine
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.5 V
p-p
into Hi-Z; 10 mV
p-p
to 1.25 V
p-p
into 50 Ω
Cardiac (typical)
Frequency range 0.1 Hz to 500 kHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Arbitrary
Memory depth 1 to 128 k
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Repetition rate 0.1 Hz to 25 MHz
Sample rate 250 MS/s