Datasheet
www.tektronix.com 29
Mixed Domain Oscilloscopes
Absolute amplitude accuracy (AAA) and channel response (CR), with TPA-N-PRE preamp attached
AAA: ≤ ±0.5 dB (typical), 18 °C –28 °C temperature range, 50 kHz to 6 GHz, either preamp state.
AAA: ≤ ±2.0 dB (typical), 18 °C –28 °C temperature range, 9 kHz to 50 kHz, either preamp state.
AAA: ≤ ±2.3 dB (typical), over full operating range, either preamp state.
CR: 0.0 dB
Crosstalk to spectrum analyzer from oscilloscope channels
≤1 GHz input frequencies
< -68 dB from ref level
>1 GHz –2 GHz input frequencies
< -48 dB from ref level
Phase noise at 1 GHz CW
1 kHz
< -104 dBc/Hz, (typical)
10 kHz
< -108 dBc/Hz, < -111 dBc/Hz (typical)
100 kHz
< -110 dBc/Hz, < -113 dBc/Hz (typical)
1 MHz
< -120 dBc/Hz, < -123 dBc/Hz (typical)
Reference frequency error (cumulative)
Cumulative error: 1.6 × 10
-6
Includes allowances for aging per year, reference frequency calibration accuracy, and temperature stability Valid
over the recommended one-year calibration interval, from 0 °C to +50 °C
Marker frequency measurement accuracy
±((1.6 × 10
-6
× Marker-Frequency) + (0.001 × span + 2)) Hz
Example: assuming the span is set to 10 kHz and the marker is at 1500 MHz, this would result in a Frequency
Measurement Accuracy of +/-((1.6 × 10
-6
× 1500 MHz) + (0.001 × 10 kHz + 2)) = ±2.412 kHz.
Marker Frequency with Span/RBW ≤ 1000:1
Reference Frequency Error with Marker level to displayed noise level > 30 dB
Frequency measurement resolution
1 Hz
Maximum operating input level
Average continuous power
+30 dBm (1 W) for reference levels ≥ -20 dBm
+24 dBm (0.25 W) for reference levels < -20 dBm
DC maximum before damage
±40 VDC
Maximum power before
damage (CW)
+32 dBm (1.6 W) for reference levels ≥ -20 dBm
+25 dBm (0.32 W) for reference levels of < -20 dBm
Maximum power before
damage (pulse)
Peak Pulse Power: +45 dBm (32 W)
Peak Pulse Power defined as <10 µs pulse width, <1% duty cycle, and
reference level of ≥ +10 dBm
Maximum operating input level with TPA-N-PRE preamp attached
Average continuous power
+30 dBm (1 W)
DC maximum before damage
±20 VDC
Maximum power before
damage (CW)
+30 dBm (1 W)
Maximum power before
damage (pulse)
+45 dBm (32 W) (<10 µs pulse width, <1% duty cycle,
and reference level of ≥ +10 dBm)