Datasheet

Datasheet
www.tektronix.com24
Trigger types
Edge Positive, negative, or either slope on any channel. Coupling includes DC, AC, HF reject, LF reject, and
noise reject.
Sequence (B-trigger) Trigger Delay by Time: 8 ns to 8 s. Or Trigger Delay by Events: 1 to 4, 000, 000 events. Not available when
“Either” edge is selected.
Pulse Width Trigger on width of positive or negative pulses that are >, <, =, , or inside/outside a specified period of time.
Timeout Trigger on an event which remains high, low, or either, for a specified time period (4 ns to 8 s).
Runt Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.
Logic Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can
be used as a clock to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all input
channels defined as High, Low, or Don’t Care.
Setup and Hold Trigger on violations of both setup time and hold time between clock and data present on any of the analog and
digital input channels.
Setup and hold trigger type Description
Setup Time Range -0.5 ns to 1.024 ms
Hold Time Range 1.0 ns to 1.024 ms
Setup + Hold Time Range 0.5 ns to 2.048 ms
Rise/Fall Time Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either and
time range is 4.0 ns to 8 s.
Video Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals.
Custom bi-level and tri-level sync video standards.
Extended Video (optional) Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF,
1080p/25, 1080p/30, 1080p/50, 1080p/60, and custom bi-level and tri-level sync video standards.
Custom bi-level and tri-level sync video standards.
Parallel (requires option MDO4MSO)
Trigger on a parallel bus data value. Parallel bus can be from 1 to 20 bits (from the digital and analog channels) in
size. Binary and Hex radices are supported.
Acquisition system
Acquisition modes
Sample Acquire sampled values.
Peak Detect Captures glitches as narrow as 800 ps (MDO4104C with option SA3 or SA6 and 2 channels enabled or
MDO4104C without SA3 or SA6) or 1.6 ns MDO4104C with option SA3 or SA6 and 3 channels enabled and all
other models) at all sweep speeds
Averaging From 2 to 512 waveforms included in average.
Envelope Min-max envelope reflecting Peak Detect data over multiple acquisitions. Number of waveforms in the envelope
selectable between 1 and 2000 and infinity
Hi Res Real-time boxcar averaging reduces random noise and increases vertical resolution.
Roll Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div.
FastAcq
®
FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events, capturing
>340, 000 wfms/son 1 GHz models and >270,000 wfms/s on 200 MHz – 500 MHz models.