www.tektronix.com 23
Mixed Domain Oscilloscopes
Random noise
Vertical scale setting
50 Ω, RMS
MDO4104C
(All congurations)
MDO40x4C (with
option SA3 or SA6)
MDO40x4 (without
option SA3 or SA6)
1 mV/div 0.093 mV 0.084 mV 0.163 mV
100 mV/div 3.31 mV 2.37 mV 2.01 mV
1 V/div 24.27 mV 20.62 mV 20.51 mV
Oset range
Volts/div setting
Oset range
1 M Ω input 50 Ω input
1 mV/div to 50 mV/div ±1 V ±1 V
50.5 mV/div to 99.5 mV/div ±0.5 V ±0.5 V
100 mV/div to 500 mV/div ±10 V ±10 V
505 mV/div to 995 mV/div ±5 V ±5 V
1 V/div to 10 V/div ±100 V ±5 V
5.05 V/div to 10 V/div ±50 V NA
Horizontal system analog channels
Time base range
1 GHz models (without option SA3 or SA6) and 1 GHz models (with option SA3 or SA6 with 2 channels enabled)
400 ps to 1000 s
≤ 500 MHz models and 1 GHz models (with option SA3 or SA6 and 4 channels enabled)
1 ns to 1000 s
Maximum duration at highest sample rate (all/half channels)
1 GHz models (without option SA3 or SA6) and 1 GHz models (with option SA3 or SA6 with 2 channels enabled)
8/4 ms
≤ 500 MHz models and 1 GHz models (with option SA3 or SA6 and 4 channels enabled)
8/8 ms
Time-base delay time range -10 divisions to 5000 s
Channel-to-channel deskew range
±125 ns
Time base accuracy ±5 ppm over any ≥1 ms interval
Trigger system
Trigger modes Auto, Normal, and Single
Trigger coupling DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
Trigger holdo range 20 ns to 8 s
Trigger sensitivity Internal DC coupled
1 mV/div to 4.98 mV/div 1.8 div
5 mV/div to 9.98 mV/div 0.6 div
10 mV/div to 19.98 mV/div 1.2 div
≤20 mV/div 0.5 div
Trigger level ranges
Any input channel ±8 divisions from center of screen, ±8 divisions from 0 V when vertical LF reject trigger coupling is selected
Line The line trigger level is fixed at about 50% of the line voltage.
Trigger frequency readout Provides 6-digit frequency readout of triggerable events.