Datasheet
Datasheet
www.tektronix.com22
Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
1. Oscilloscope
MDO4024C MDO4034C MDO4054C MDO4104C
Analog channels 4 4 4 4
Analog channel bandwidth 200 MHz 350 MHz 500 MHz 1 GHz
Rise time 1.75 ns 1 ns 700 ps 350 ps
Sample rate (1 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s
Sample rate (2 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s
Sample rate (4 ch)
Without option SA3 or SA6
With option SA3 or SA6
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
2.5 GS/s
5 GS/s
2.5 GS/s
Record length (1 ch) 20 M 20 M 20 M 20 M
Record length (2 ch) 20 M 20 M 20 M 20 M
Record length (4 ch) 20 M 20 M 20 M 20 M
Digital channels with
MDO4MSO option
16 16 16 16
Arbitrary Function Generator
outputs with MDO4AFG option
1 1 1 1
Spectrum analyzer channels with
option SA3 or SA6
1 1 1 1
Spectrum analyzer
frequency range
With Option SA3
With Option SA6
9 kHz – 3 GHz
9 kHz – 6 GHz
9 kHz – 3 GHz
9 kHz – 6 GHz
9 kHz – 3 GHz
9 kHz – 6 GHz
9 kHz – 3 GHz
9 kHz – 6 GHz
Vertical System Analog Channels
Hardware bandwidth limits
≥350 MHz models 20 MHz or 250 MHz
200 MHz models 20 MHz
Input coupling AC, DC
Input impedance 1 MΩ ±1% (13 pF), 50 Ω ±1%
Input sensitivity range
1 MΩ 1 mV/div to 10 V/div
50 Ω 1 mV/div to 1 V/div
Vertical resolution 8 bits (11 bits with Hi Res)
Maximum input voltage
1 MΩ 300 VRMS CAT II with peaks ≤ ±425V
50 Ω 5 VRMS with peaks ≤ ±20V
DC gain accuracy ±1.5%, derated at 0.10%/°C above 30 °C
±3.0% for variable gain, derated 0.10%/°C above 30 °C
Oset accuracy ±(0.005 * |offset – position| + DC Balance)
DC balance 0.1 div with DC–50 Ω scope input impedance (50Ω BNC terminated)
Channel-to-channel isolation (typical)
Any two channels at equal vertical scale ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth