Mixed Domain Oscilloscopes WINNER OF 13 INDUSTRY AWARDS MDO4000C Series Datasheet Key Performance Specifications 1. Oscilloscope 4 analog channels 1 GHz, 500 MHz, 350 MHz, and 200 MHz bandwidth models Bandwidth is upgradable (up to 1 GHz) Up to 5 GS/s sample rate 20 M record length on all channels > 340, 000 wfm/s maximum waveform capture rate Standard passive voltage probes with 3.
Datasheet MSO/DPO2000B MDO4000C MSO/DPO5000B Advanced debug features at an affordable price 6-in-1 integrated oscilloscope Performance 6-in-1 integrated oscilloscope with synchronized insights into analog, digital and spectral signals Exceptional signal fidelity with advanced analysis and math Commonly Used For Design and Debug Education Design and Debug EMI Troubleshooting Education Design and Debug EMI Troubleshooting General Purpose RF Design and Integration Advanced Design and Debug USB Ether
Mixed Domain Oscilloscopes Typical Applications Embedded Design Discover and solve issues quickly by performing system level debug on mixed signal embedded systems including today’s most common serial bus and wireless technologies. 1.
Datasheet The Spectral palette uses color-grading to indicate frequency of occurrence with colder colors like blue indicating frequently occurring events and hot colors like red indicating rarely occurring events. The Normal palette uses the default channel color (like yellow for channel one) along with gray-scale to indicate frequency of occurrence where frequently occurring events are bright.
Mixed Domain Oscilloscopes Zoom and Pan A dedicated, two-tier front-panel control provides intuitive control of both zooming and panning. The inner control adjusts the zoom factor (or zoom scale); turning it clockwise activates zoom and goes to progressively higher zoom factors, while turning it counterclockwise results in lower zoom factors and eventually turning zoom off. No longer do you need to navigate through multiple menus to adjust your zoom view.
Datasheet Waveform Analysis Waveform Histograms Verifying that your prototype’s performance matches simulations and meets the project’s design goals requires analyzing its behavior. Tasks can range from simple checks of rise times and pulse widths to sophisticated power loss analysis and investigation of noise sources. show visually how waveforms vary over time.
Mixed Domain Oscilloscopes Video Design and Development (Optional) Power Analysis (Optional) Many video engineers have remained loyal to analog oscilloscopes, believing the intensity gradations on an analog display are the only way to see certain video waveform details. The fast waveform capture rate, coupled with its intensity-graded view of the signal, provides the same information-rich display as an analog oscilloscope, but with much more detail and all the benefits of digital scopes.
Datasheet Limit-Mask Testing (Optional) 2. Spectrum Analyzer (Optional) A common task during the development process is characterizing the behavior of certain signals in a system. One method, called limit testing, is to compare a tested signal to a known good or “golden” version of the same signal with user-defined vertical and horizontal tolerances. Another common method, called mask testing, is to compare a tested signal to a mask, looking for where a signal under test violates the mask.
Mixed Domain Oscilloscopes Two manual markers are also available for measuring nonpeak portions of the spectrum. When enabled, the reference marker is attached to one of the manual markers, enabling delta measurements from anywhere in the spectrum. In addition to frequency and amplitude, manual marker readouts also include noise density and phase noise readouts depending on whether Absolute or Delta readouts are selected.
Datasheet Ultra-Wide Capture Bandwidth Spectrum Traces Today’s wireless communications vary significantly with time, using sophisticated digital modulation schemes and, often, transmission techniques that involve bursting the output. These modulation schemes can have very wide bandwidth as well. Traditional swept or stepped spectrum analyzers are ill equipped to view these types of signals as they are only able to look at a small portion of the spectrum at any one time.
Mixed Domain Oscilloscopes The optional MDO4TRIG application module provides advanced RF triggering. This module enables the RF power level on the spectrum analyzer to be used as a source for Sequence, Pulse Width, Timeout, Runt, and Logic trigger types. For example, you can trigger on a RF pulse of a specific length or use the spectrum analyzer channel as an input to a logic trigger, enabling the oscilloscope to trigger only when the RF is on while other signals are active.
Datasheet Visualizing Changes in your RF Signal Advanced RF Analysis The time domain graticule on the MDO4000C Series display provides support for three RF time domain traces that are derived from the underlying I and Q data of the spectrum analyzer input including: When paired with SignalVu-PC and its Live Link option, the MDO4000C Series becomes the industry’s widest bandwidth Vector Signal Analyzer with up to 1 GHz capture bandwidth.
Mixed Domain Oscilloscopes of interest, but never both at the same time. This is because traditional oscilloscopes only have a single acquisition system with a single set of user settings such as record length, sample rate, and time per division that drive all data views. But with the MDO4000C Series, the spectrum analyzer has its own acquisition system that is independent, but time correlated, to the analog and digital channel acquisition systems.
Datasheet 3. Arbitrary Function Generator (Optional) The MDO4000C contains an optional integrated arbitrary function generator (option MDO4AFG), perfect for simulating sensor signals within a design or adding noise to signals to perform margin testing. The integrated function generator provides output of predefined waveforms up to 50 MHz for sine, square, pulse, ramp/triangle, DC, noise, sin(x)/x (Sinc), Gaussian, Lorentz, exponential rise/fall, Haversine and cardiac.
Mixed Domain Oscilloscopes 4. Logic Analyzer (Optional) The logic analyzer (option MDO4MSO) provides 16 digital channels which are tightly integrated into the oscilloscope’s user interface. This simplifies operation and makes it possible to solve mixed-signal issues easily. With color-coded digital waveform display, groups are created by simply placing digital channels together on the screen, allowing digital channels to be moved as a group.
Datasheet P6616 MSO Probe This unique probe design offers two eight-channel pods. Each channel ends with a probe tip featuring a recessed ground for simplified connection to the device under test. The coax on the first channel of each pod is colored blue making it easy to identify. The common ground uses an automotivestyle connector making it easy to create custom grounds for connecting to the device under test.
Mixed Domain Oscilloscopes Serial bus technologies supported by the MDO4000C Technology Embedded Trigger, Decode, Search Order product IC Yes DPO4EMBD SPI Yes DPO4EMBD Computer RS232/422/485, UART Yes DPO4COMP USB USB LS, FS, HS Yes (trigger on LS FS, HS) HS available only on 1 GHz models DPO4USB Ethernet 10Base-T, 100Base-TX Yes DPO4ENET Automotive CAN Yes DPO4AUTO or DPO4AUTOMAX LIN Yes DPO4AUTO or DPO4AUTOMAX FlexRay Yes DPO4AUTOMAX MIL-STD-1553 Yes DPO4AERO Military an
Datasheet 6. Digital Voltmeter (DVM) and Frequency Counter The MDO4000C Series Platform The MDO4000C contains an integrated 4-digit digital voltmeter (DVM) and 5-digit frequency counter. Any of the analog inputs can be a source for the voltmeter, using the same probes that are already attached for general oscilloscope usage. The easy-to-read display offers you both numeric and graphical representations of the changing measurement values.
Mixed Domain Oscilloscopes Remote Connectivity and Instrument Control Probing Exporting data and measurements is as simple as connecting a USB cable from the oscilloscope to your PC. Key software applications – OpenChoice® Desktop, and Microsoft Excel and Word toolbars – are included standard with each oscilloscope to enable fast and easy direct communication with your Windows PC. The MDO4000C Series scope ships standard with passive voltage probes and uses the TekVPI probe interface.
Datasheet The Ultimate 6-in-1Integrated Oscilloscope,Completely C 1. Oscilloscope 2. Spectrum Analyzer 3. Arbitrary Function Generator 4. Logic Analyzer 5. Protocol Analyzer 6. Digital Voltmeter & Frequency Counter Mixed domain oscilloscope Fast acquisition, selectable record lengths, and a full suite of automated measurements help solve debug challenges quickly Large 10.
Mixed Domain Oscilloscopes Customizable and Fully Upgradable 5 Serial Protocol Analysis and Application modules add triggering and decode on common serial protocols, automated power measurements, and limit/mask test Wave Inspector navigation and search simplifies finding areas of interest in deep records.
Datasheet Specifications All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise. 1. Oscilloscope MDO4024C MDO4034C MDO4054C MDO4104C 4 4 4 4 200 MHz 350 MHz 500 MHz 1 GHz 1.75 ns 1 ns 700 ps 350 ps Sample rate (1 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s Sample rate (2 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s Sample rate (4 ch) Without option SA3 or SA6 With option SA3 or SA6 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.
Mixed Domain Oscilloscopes Random noise 50 Ω, RMS Vertical scale setting 1 mV/div MDO4104C (All configurations) MDO40x4C (with option SA3 or SA6) MDO40x4 (without option SA3 or SA6) 0.093 mV 0.084 mV 0.163 mV 100 mV/div 3.31 mV 2.37 mV 2.01 mV 1 V/div 24.27 mV 20.62 mV 20.51 mV Offset range Offset range Volts/div setting 1 mV/div to 50 mV/div 1 M Ω input 50 Ω input ±1 V ±1 V 50.5 mV/div to 99.5 mV/div ±0.5 V ±0.
Datasheet Trigger types Edge Positive, negative, or either slope on any channel. Coupling includes DC, AC, HF reject, LF reject, and noise reject. Sequence (B-trigger) Trigger Delay by Time: 8 ns to 8 s. Or Trigger Delay by Events: 1 to 4, 000, 000 events. Not available when “Either” edge is selected. Pulse Width Trigger on width of positive or negative pulses that are >, <, =, ≠, or inside/outside a specified period of time.
Mixed Domain Oscilloscopes Waveform measurements Cursors Waveform and Screen DC measurement accuracy ±((DC Gain Accuracy) * |reading – (Offset – Position)| + Offset Accuracy + 0.15 div + 0.6 mV) Automatic measurements (time domain) 30, of which up to eight can be displayed on-screen at any one time.
Datasheet Power measurements (optional, requires DPO4PWR) Power quality measurements VRMS, VCrestFactor, Frequency, IRMS, ICrestFactor, TruePower, ApparentPower, ReactivePower, PowerFactor, PhaseAngle. Switching loss measurements Power loss Ton, Toff, Conduction, Total. Energy loss Ton, Toff, Conduction, Total. Harmonics THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MIL-STD-1399, Section 300A. Ripple measurements VRippleandIRipple.
Mixed Domain Oscilloscopes 2.
Datasheet Spurious response 2nd and 3rd harmonic distortion (>100 MHz) < -60 dBc (< -65 dBc, typical), with auto settings on and signals 10 dB below reference level 2nd and 3rd harmonic distortion (9 kHz to 100 MHz ) < -57 dBc (< -60 dBc, typical), with auto settings on, signals 10 dB below reference level, and reference level ≤ -15 dBm 2nd order intermodulation distortion (>200 MHz) < -60 dBc (< -65 dBc, typical), with auto settings on and signals 10 dB below reference level 2nd order intermodulation dist
Mixed Domain Oscilloscopes Absolute amplitude accuracy (AAA) and channel response (CR), with TPA-N-PRE preamp attached AAA: ≤ ±0.5 dB (typical), 18 °C –28 °C temperature range, 50 kHz to 6 GHz, either preamp state. AAA: ≤ ±2.0 dB (typical), 18 °C –28 °C temperature range, 9 kHz to 50 kHz, either preamp state. AAA: ≤ ±2.3 dB (typical), over full operating range, either preamp state. CR: 0.
Datasheet RF Power level trigger Frequency range Models with option SA3: 1 MHz to 3 GHz Models with option SA6: 1 MHz to 3.75 GHz; 2.75 GHz to 4.5 GHz, 3.5 GHz to 6.
Mixed Domain Oscilloscopes Square/Pulse Frequency range 0.1 Hz to 25 MHz Amplitude range 20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω Duty cycle 10% to 90% or 10 ns minimum pulse, whichever is larger cycle Duty cycle resolution 0.
Datasheet Frequency accuracy Sine wave and ramp 130 ppm (frequency < 10 kHz) 50 ppm (frequency ≥ 10 kHz) Square wave and pulse 130 ppm (frequency < 10 kHz) 50 ppm (frequency ≥ 10 kHz) Resolution 0.1 Hz or 4 digits, whichever is larger Amplitude accuracy ±[ (1.5% of peak-to-peak amplitude setting) + (1.5% of DC offset setting) + 1 mV ] (frequency = 1 kHz) DC offset DC offset range ±2.5 V into Hi-Z; ±1.25 V into 50 Ω DC offset resolution 1 mV into Hi-Z; 500 uV into 50 Ω Offset accuracy ±[(1.
Mixed Domain Oscilloscopes 5. Serial Protocol Analyzer (optional) AutomatedSerial Triggering, Decode, and Search options for I2C, SPI, RS-232/422/485/UART, USB2.0, CAN, LIN, FlexRay, MIL-STD-1553, and Audio buses. For more detailed information about serial bus support products please see the Serial Triggering and Analysis Application Modules data sheet. Trigger types I2C Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 10 Mb/s.
Datasheet Ethernet3 10BASE-T and 100BASE-TX: Trigger on Start Frame Delimiter, MAC Addresses, MAC Q-Tag Control Information, MAC Length/ Type, IP Header, TCP Header, TCP/IPv4/MAC Client Data, End of Packet, and FCS (CRC) Error. 100BASE-TX: Idle. MAC Addresses – Trigger on Source and Destination 48-bit address values. MAC Q-Tag Control Information – Trigger on Q-Tag 32-bit value. MAC Length/Type – Trigger on ≤, <, =, >, ≥, ≠ a particular 16-bit value, or inside or outside of a range.
Mixed Domain Oscilloscopes General Product Specifications Display system Display type 10.4 in.
Datasheet e*Scope® Web-based interface Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or network name of the oscilloscope and a web page will be served to the browser. Transfer and save settings, waveforms, measurements, and screen images or make live control changes to settings on the oscilloscope directly from the web browser.
Mixed Domain Oscilloscopes Ordering information Step 1 Choose the MDO4000C base model MDO4000C family MDO4024C Mixed Domain Oscilloscope with (4) 200 MHz analog channels MDO4034C Mixed Domain Oscilloscope with (4) 350 MHz analog channels MDO4054C Mixed Domain Oscilloscope with (4) 500 MHz analog channels MDO4104C Mixed Domain Oscilloscope with (4) 1 GHz analog channels Standard accessories Probes ≤ 500 MHz models TPP0500B, 500 MHz bandwidth, 10X, 3.9 pF.
Datasheet Power cord and plug options Opt. A0 North America power plug (115 V, 60 Hz) Opt. A1 Universal Euro power plug (220 V, 50 Hz) Opt. A2 United Kingdom power plug (240 V, 50 Hz) Opt. A3 Australia power plug (240 V, 50 Hz) Opt. A5 Switzerland power plug (220 V, 50 Hz) Opt. A6 Japan power plug (100 V, 50/60 Hz) Opt. A10 China power plug (50 Hz) Opt. A11 India power plug (50 Hz) Opt. A12 Brazil power plug (60 Hz) Opt.
Mixed Domain Oscilloscopes Step 3 Select application modules and accessories Application modules Application modules are purchased as stand-alone products and can be purchased at the time of initial MDO4000C purchase or at any future time. The optional application modules functionality is offered free for a 30-day trial period. This free trial period starts automatically when the instrument is powered on for the first time.
Datasheet DPO4PWR Power Analysis Application Module. Enables quick and accurate analysis of power quality, switching loss, harmonics, safeoperating area (SOA), modulation, ripple, and slew rate (dI/dt, dV/dt). DPO4LMT Limit and Mask Testing Application Module. Enables testing against limit templates generated from “golden” waveforms and mask testing using custom masks.6 DPO4VID HDTV and Custom (nonstandard) Video Triggering and Video Picture Module.
Mixed Domain Oscilloscopes Other RF probes Contact Beehive Electronics to order: http://beehive-electronics.com/probes.html 101A EMC probe set 150A EMC probe amplifier 110A Probe cable 0309-0001 SMA probe adapter 0309-0006 BNC probe adapter Step 4 Add instrument upgrades in the future Instrument upgrades The MDO4000C Series products offer a number of ways to add functionality after the initial purchase.
Datasheet Post-purchase service products The following upgrades can be added to any model to extend the product warranty beyond the standard warranty period. MDO4024C-R5DW Repair service coverage 5 years for a MDO4024C product (includes product warranty period). MDO4034C-R5DW Repair service coverage 5 years for a MDO4034C product (includes product warranty period). MDO4054C-R5DW Repair service coverage 5 years for a MDO4054C product (includes product warranty period).
Mixed Domain Oscilloscopes www.tektronix.
Contact Information: ASEAN / Australia (65) 6356 3900 Austria 00800 2255 4835 Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium 00800 2255 4835 Brazil +55 (11) 3759 7627 Canada 1 800 833 9200 Central East Europe and the Baltics +41 52 675 3777 Central Europe & Greece +41 52 675 3777 Denmark +45 80 88 1401 Finland +41 52 675 3777 France 00800 2255 4835 Germany 00800 2255 4835 Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835 Japan 81 (3) 6714 3010 Luxembourg +41