Datasheet
LAN eXtensions for Instrumentation (LXI)
Class LXI Core 2011
Version V1.4
Power source
Power source voltage 100 to 240 V ±10%
Power source frequency 50 to 60 Hz at 100 to 240 V
400 Hz ±10% at 115 V
Power consumption 120 W maximum
Physical characteristics
Dimensions
Height 203.2 mm (8 in.)
Width 416.6 mm (16.4 in.)
Depth 147.4 mm (5.8 in.)
Weight
Net 4.2 kg (9.2 lb.)
Shipping 8.6 kg (19 lb.)
Rackmount configuration 5U
Cooling clearance 2 in. (51 mm) required on left side and rear of instrument
EMC, environment, and safety
Temperature
Operating -10 ºC to +55 ºC (+14 ºF to 131 ºF)
Nonoperating -40 ºC to +71 ºC (-40 ºF to 160 ºF)
Humidity
Operating Up to +40 ºC, 5% to 90% relative humidity
+40 ºC to +55 ºC, 5% to 60% relative humidity
Nonoperating Up to +40 ºC, 5% to 90% relative humidity
Above +40 ºC up to +55 ºC, 5% to 60% relative humidity
Above +55 ºC up to +71 ºC, 5% to 40% relative humidity, non-condensing
Altitude
Operating 3,000 meters (9,843 feet)
Nonoperating 12,000 meters (39,370 feet)
Regulatory
Electromagnetic compatibility EC Council Directive 2004/108/EC
Safety UL61010-1:2004, CAN/CSA-C22.2 No. 61010.1: 2004, Low Voltage Directive 2006/95/EC and EN61010-1:2001, IEC
61010-1:2001, ANSI 61010-1-2004, ISA 82.02.01
Mixed Domain Osciloscopes -- MDO3000 Series
www.tektronix.com 27