Datasheet
Amplitude accuracy ±[ (1.5% of peak-to-peak amplitude setting) + (1.5% of DC offset setting) + 1 mV ] (frequency = 1 kHz)
DC offset
DC offset range ±[2.5 V – (signal amplitude) / 2 ] into Hi-Z; ±[1.25 – (signal amplitude) / 2 ] into 50 Ω
DC offset resolution 1 mV into Hi-Z; 500 uV into 50 Ω
Offset accuracy ±[(1.5% of absolute offset voltage setting) + 1 mV]; derated 3 mV for every 10 °C away from 25 °C
ArbExpress
®
The MDO3000 is compatible with ArbExpress
®
PC-based signal generator waveform creation and editing software. Capture
waveforms on the MDO3000 oscilloscope and transfer them to ArbExpress for editing. Create complex waveforms in ArbExpress
and transfer them to the arbitrary function generator in the MDO3000 for output. To download ArbExpress software, go to
www.tektronix.com/downloads.
Digital voltmeter (DVM) and frequency counter
Source Channel 1, Channel 2, Channel 3, Channel 4
Measurement types AC RMS, DC, AC+DC RMS (reads out in volts or amps); Frequency
Resolution ACV, DCV: 4 digits
Frequency: 5 digits
Frequency accuracy 10 ppm
Measuring rate 100 times/second; measurements updated on the display 4 times/second
Vertical settings autorange Automatic adjustment of vertical settings to maximize measurement dynamic range; available for any non-trigger source
Graphical measurement Graphical indication of minimum, maximum, current value, and five second rolling range
Software
OpenChoice
®
Desktop Enables fast and easy communication between a Windows PC and your oscilloscope using USB or LAN. Transfer and save
settings, waveforms, measurements, and screen images. Word and Excel toolbars automate the transfer of acquisition data and
screen images from the oscilloscope into Word and Excel for quick reporting or further analysis.
IVI driver Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI,
MicrosoftNET, and MATLAB.
e*Scope
®
Web-based interface Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or
network name of the oscilloscope and a web page will be served to the browser. Transfer and save settings, waveforms,
measurements, and screen images or make live control changes to settings on the oscilloscope directly from the web browser.
LXI Core 2011 Web interface Connect to the oscilloscope through a standard Web browser by simply entering the oscilloscope IP address or network name in
the address bar of the browser. The Web interface enables viewing of instrument status and configuration, status and modification
of network settings, and instrument control through e*Scope Web-based remote control. All Web interaction conforms to LXI Core
2011 specification, version 1.4.
Mixed Domain Osciloscopes -- MDO3000 Series
Arbitrary Function Generator
www.tektronix.com 25