Datasheet

Limit/Mask testing (optional)
Test source Limit test: Any Ch1 - Ch4 or any R1 - R4
Mask test: Any Ch1 - Ch4
Mask creation Limit test vertical tolerance from 0 to 1 division in 1 m division increments; Limit test horizontal tolerance from 0 to 500 m division in
1 m division increments.
Load custom mask from text file with up to 8 segments.
Mask scaling Lock to Source ON (mask automatically re-scales with source-channel settings changes)
Lock to Source OFF (mask does not re-scale with source-channel settings changes)
Test criteria run until Minimum number of waveforms (from 1 to 1,000,000 and Infinity)
Minimum elapsed time (from 1 second to 48 hours and Infinity)
Violation threshold From 1 to 1,000,000 and Infinity
Actions on test failure Stop acquisition, save screen image to file, save waveform to file, print screen image, AUX OUT pulse, set remote interface SRQ
Actions on test complete AUX OUT pulse, set remote interface SRQ
Results display Test status, total waveforms, number of violations, total tests, failed tests, elapsed time, total hits for each mask segment
Arbitrary Function Generator
(Requires MDO3AFG option)
Waveforms Sine, Square, Pulse, Ramp/Triangle, DC, Noise, Sin(x)/x (Sinc), Gaussian, Lorentz, Exponential Rise, Exponential Decay,
Haversine, Cardiac, and Arbitrary.
Sine
Frequency range 0.1 Hz to 50 MHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Amplitude flatness ±0.5 dB typical at 1 kHz (±1.5 dB for <20 mV
p-p
amplitudes)
Total harmonic distortion
(typical)
1% into 50Ω
2% for amplitude < 50 mV and frequencies > 10 MHz
3% for amplitude < 20 mV and frequencies > 10 MHz
Spurious free dynamic range
(SFDR)
-40 dBc (V
p-p
≥ 0.1 V); -30dBc (V
p-p
≤ 0.1 V), 50 Ω load
Square / Pulse
Frequency range 0.1 Hz to 25 MHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Duty cycle 10% to 90% or 10 ns minimum pulse, whichever is larger cycle
Duty cycle resolution 0.1%
Pulse width minimum 10 ns typical
Rise/fall time 5 ns typical (10% - 90%)
Pulse width resolution 100 ps
Overshoot < 2% typical for signal steps greater than 100 mV
Asymmetry ±1% ±5 ns, at 50% duty cycle
Jitter (TIE RMS) < 500 ps typical
Mixed Domain Osciloscopes -- MDO3000 Series
www.tektronix.com 23