Datasheet

Horizontal system analog channels
Time base range
1 GHz models 400 ps/div to 1000 s/div
≤ 500 MHz models 1 ns/div to 1000 s/div
Maximum duration at highest
sample rate (all/half channels)
1 GHz models 4/2 ms
≤ 500 MHz models 4/4 ms
Time-base delay time range -10 divisions to 5000 s
Channel-to-channel deskew range ±125 ns
Time base accuracy ±10 ppm over any ≥1 ms interval
Horizontal system digital channels
(Requires MDO3MSO option)
Maximum sample rate (Main) 500 MS/s (2 ns resolution)
Maximum record length (Main) 10 M
Maximum sample rate (MagniVu) 8.25 GS/s (121.2 ps resolution)
Maximum record length (MagniVu 10k centered on the trigger
Minimum detectable pulse width
(typical)
2 ns
Channel-to-channel skew (typical) 500 ps
Maximum input toggle rate 250 MHz (Maximum frequency sine wave that can accurately be reproduced as a logic square wave. Requires the use of a short
ground extender on each channel. This is the maximum frequency at the minimum swing amplitude. Higher toggle rates can be
achieved with higher amplitudes.)
Spectrum analyzer input
Capture bandwidth MDO3012, MDO3014 models: 100 MHz
MDO3022, MDO3024 models: 200 MHz
MDO3032, MDO3034 models: 350 MHz
MDO3052, MDO3054 models: 500 MHz
MDO3102, MDO3104 models: 1 GHz
All models: 3 GHz with option MDO3SA
Span MDO3012, MDO3014 models: 9 kHz – 100 MHz
MDO3022, MDO3024 models: 9 kHz – 200 MHz
MDO3032, MDO3034 models: 9 kHz – 350 MHz
MDO3052, MDO3054 models: 9 kHz – 500 MHz
MDO3102, MDO3104 models: 9 kHz – 1 GHz
All models: 9 kHz – 3 GHz with option MDO3SA, in a 1-2-5 sequence
Datasheet
16 www.tektronix.com