Datasheet

Specifications
All specifications apply to all models unless noted otherwise.
MDO3012 MDO3014 MDO3022 MDO3024 MDO3032 MDO3034 MDO3052 MDO3054 MDO3102 MDO3104
Analog channels 2 4 2 4 2 4 2 4 2 4
Analog channel bandwidth 100 MHz 100 MHz 200 MHz 200 MHz 350 MHz 350 MHz 500 MHz 500 MHz 1 GHz 1 GHz
Rise time
(10 mV/div setting with
50Ω input termination)
4 ns 4 ns 2 ns 2 ns 1.14 ns 1.14 ns 800 ps 800 ps 400 ps 400 ps
Sample rate (1 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s
Sample rate (2 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s
Sample rate (4 ch) - 2.5 GS/s - 2.5 GS/s - 2.5 GS/s - 2.5 GS/s - 2.5 GS/s
Record length (1 ch) 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M
Record length (2 ch) 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M 10 M
Record length (4 ch) - 10 M - 10 M - 10 M - 10 M - 10 M
Digital channels with
MDO3MSO option
16 16 16 16 16 16 16 16 16 16
Arbitrary Function
Generator outputs with
MDO3AFG option
1 1 1 1 1 1 1 1 1 1
Spectrum analyzer
channels
1 1 1 1 1 1 1 1 1 1
Standard spectrum
analyzer frequency range
9 kHz -
100 MHz
9 kHz -
100 MHz
9 kHz -
200 MHz
9 kHz -
200 MHz
9 kHz -
350 MHz
9 kHz -
350 MHz
9 kHz -
500 MHz
9 kHz -
500 MHz
9 kHz -
1 GHz
9 kHz -
1 GHz
Optional spectrum
analyzer frequency range
with MDO3SA option
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
9 kHz -
3 GHz
Vertical system analog channels
Hardware bandwidth limits
≥350 MHz models 20 MHz or 250 MHz
100 MHz and 200 MHz models 20 MHz
Input coupling AC, DC
Input impedance 1 MΩ ±1%, 50 Ω ±1%, 75 Ω ±1%; 75 Ω not available on 1 GHz models
Input sensitivity range
1 MΩ 1 mV/div to 10 V/div
50 Ω, 75 Ω 1 mV/div to 1 V/div
Vertical resolution 8 bits (11 bits with Hi Res)
Maximum input voltage
1 MΩ 300 V
RMS
CAT II with peaks ≤ ±425 V
50 Ω, 75 Ω 5 V
RMS
with peaks ≤ ±20 V
Datasheet
14 www.tektronix.com