User manual
Compliance information
This section lists the EMC (electromagnetic compliance), s afety, and environmental standards with which the instrument
complies.
EMC compliance
EC Declaration of Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following
specifications as listed in the Official Journal of the European Communities:
EN 61326-1
:2006, EN 61326-2-1:2006.
EMC requirements for electrical equipment for measurement, control, and
laboratory use.
123
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic field immunity
4
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
5
IEC 61000-4-11 :2004. Voltage dips and interruptions immunity
6
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, fluctuations, and flicker
European contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the l evels required by this standard may occur when this equipment is connected to a test object.
3
For c
ompliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Oscilloscope: ≤ 3.0 division waveform displacement and ≤ 6.0 division increase in peak-to-peak noise. RF: Residual spurious
sign
als in the RF section can typically increase to –50 dBm when the instrument is subjected to electromagnetic interference per the
IEC 61000-4-3 test for frequencies up to 1 GHz, and to –35 dBm for frequencies above 1 GHz.
5
Osci
lloscope: ≤ 1.0 division waveform displacement and ≤ 2.0 division increase in pe ak-to-peak noise. RF: Residual spurious
signals in the RF section can typically increase to –85 dBm when the instrument is subjected to electromagnetic interference
per the IEC 61000-4-6 test.
6
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC
61000-4-11).