Trigger Setup
Using Sequence Trigger (A (Main) and B (Delayed))
Combine an edge A Event (Main) trigger with the B Event (Delayed) trigger to capture m ore complex signals. After the A
Event occurs, the trigger system looks for the B Event before triggering and displaying the waveform.
A and B triggers can (and typically do) have separate sources.
NOTE. You can select sequence triggering when you choose the slope type Falling or Rising but not when you pick
the slope type Both.
Use the Edge trigger menu to set up the A trigger rst. Then, to use the B trigger:
1. P ush Trigger Menu.
2. Push Type.
3. Turn Multipurpose a to select a trigger type
of Sequence (B Trigger).
This brings up the Sequence (B Trigger)
menu.
4. Push BTriggerAfterA.
Type
Sequence
(B Trigger)
Source
1
Coupling
DC
Slope
Level
0.00 V
B Trigger
After A
Time
Mode
Auto
& Holdoff
98 MDO3000 Series Oscilloscopes User Manual