Datasheet
Application Note
www.tektronix.com/bench12
Figure 24 shows the MSO triggered on the rising edge of the 
input to the ADC. The unique front panel Wave Inspector
®
navigation controls are used to zoom to 20 X magnification 
around the trigger, allowing the parallel bus decode values 
to be easily seen. The ADC data is stable at the falling clock 
edge and the MSO decodes the bus values at the falling edge 
of the clock. Therefore, the parallel bus updates at the falling 
clock edge when the ADC data is stable.
The MSO’s powerful triggering finds signal faults and triggers 
on parallel or serial bus content to focus the acquisition on 
problem areas. However, once the data is acquired, triggering 
doesn’t apply any more. Manually searching through long 
record lengths can be a frustrating and time-consuming 
process. A 10 Mpoint waveform record is over 10,000 full 
resolution screens of data. It takes over two hours and forty 
five minutes to scroll through a 10 Mpoint waveform at the rate 
of one full resolution screen per second. Whereas using Wave 
Inspector to search and mark a 10 Mpoint record 6-bit data 
bus acquisition takes about 30 seconds. Once the data has 
been found and marked, navigating between occurrences is 
as simple as pressing the front-panel Previous and Next arrow 
buttons. Searches are also available for trigger types such 
as edges, pulse widths, runt, setup/hold times, logic, rise/fall 
times and bus data values.
Figure 25 shows Wave Inspector searched the ADC parallel 
bus for 00 hex value which should be at each test ramp signal 
valley. But, no 00 hex values are found as indicated by the lack 
of white triangle search marks at the top of the display and as 
indicated by the search events readout showing zero at the 
bottom of the display. No 00 hex values means that the 
ADC did not see the analog input voltage that correspond 
to a 00 hex value. The acquisition system analog signal 
conditioning circuit did not correctly process the minimum 
peaks of the test ramp signal to match the minimum ADC 
input voltage in order for the ADC to produce a 00 hex output.
Figure 25. Wave Inspector search finds no 00 hex values at test signal valleys. Figure 24. The MSO triggered on the rising edge of the input to the ADC and Wave 
Inspector zooms in so that parallel bus decode hex values are easily seen. The test 
ramp signal is on channel one. At the bottom of the display is the ADC clock on digital 
channel zero. The ADC digital output bus signals one through six are above the clock 
waveform. The ADC digital signals are grouped together into the clocked parallel bus at 
the center of the display. 










