Datasheet
Data Sheet
Rear Panel Inputs and Outputs
Characteristic
Description
Reference Input 1, 5, or 10 MHz; 0.1 to 5 V
RMS
sine; impedance ≥1kΩ
Reference Output 10 MHz; >1 V
RMS
sine into 50 Ω
Arming Input
Arming of all measuring functions
Impedance
Approx. 1 kΩ
Frequency range
DC to 80 MHz
Auxiliary Functions
Trigger Hold
off
Characteristic
Description
TimeDelayRange 20nsto2s,10nsresolution
External Start and Stop Arming
Arming can be used to synchronize the frequency and power measurements with
the start of a burst signal. Minimum burst length must exceed 100 µs.
Characteri
stic
Description
Modes
Start and Stop Arming
Input Channels A, B, or E (Ext. arming input)
Max Rep. Rate for Arming Signal
Channel A, B
160 MHz
Channel E
80 MHz
Start-time Delay Range
20 ns to 2 s, 10 ns resolution
Statistics
Characteristic
Descript
ion
Function
s
Maximum, Minimum, Mean, ΔMax-Min, Standard
Deviation, and Allan Deviation
Display Numeric, histograms, or trend plots
Sample Size
2to2×10
9
samples
Limit Qualifier Off, or capture values above, below, inside, or outside
limits
Measurement Pacing Pacing Time Range: 4 µs to 500 s
Mathe
matics
Characteristic
Description
Functions
(K*X+L)/M, (K/X+L)/M, or X/M-1. X is current reading
and K, L, and M are constants; set using the keyboard or
as frozen reference value (X
0
)
Other Functions
Characteristic
Description
Measuring Time
20 ns to 1000 s for frequency, burst, and period average.
Single cycle for other measuring functions
Time-base Reference
Internal, external, or automatic
Display Hold Freezes the result, until a new measurement is initiated
through a restart
Limit Alarm
Graphical indication on front panel and/or SRQ through
GPIB
Limit Values Lower limit, upper limit
Settings Off, or alarm if value is above, below, inside, or outside
limits
On Alarm Stop or Continue
Display
Numeric + Graphic
Stored Instrument
Setups
20. Instrument setups can be saved/recalled from
internal nonvolatile memory. 10 can be user protected
Display
Backlit LCD graphics screen for menu control, numerical
readout, and status information
Number of digits
14 digits in Numerical mode
Resolution 320 × 97 pixels
GPIB Interface
Characteristic
Description
Compatibility IEEE 488.2-1987, SCPI 199953131A Compatibility
mode
Interface Functions SH1, AH1, T6, L4, SR1, RL1, DC1, DT1, E2
Max Measurement Rate
GPIB 5k readings/s (Block mode)
500 readings/s (individual GET triggered)
To internal memory
250k readings/s
Internal Memory Size
750k readings
USB Interface
Characteristic
Description
USB Version 2.0 full speed (11 Mb/s)
Calibration
Characteristic
Description
Mode
Closed case, menu controlled
Calibration Frequencies
0.1, 1, 5, 10, 1.544, and 2.048 MHz
General Specifications
Environmental Data
Characteristic
Description
Class MIL-PRF-28800F, Class 3
Operating Temp 0 °C to +50 °C
Storage Temp –40 °C to +71 °C
Humidity
5-95% (10-30 °C)
5-75% (30-40 °C)
5-45% (40-50 °C)
Altitude
Operating: 2,000 m
Storage: 12,000 m
Safety Directive 2006/95/EC, EN61010-1, UL61010-1,
CAN/CSA C22.2 No. 61010-1
EMC EU Directive 2004/108/EC, EN61326-1, EN61326-2-1,
Class A
6 www. tektronix.com