User's Manual
#!# $%$
#!# #%!
4Ć34
This procedure shows the setups and lists the steps to check the Sweep
Rate Accuracy.
%&" %! ( '" % &#)
High
Frequency
Signal
Generator
W
W
Output
#!&# %! ( '" % &#)
+%" +Install the SDĆ24 Sampling Head into the compartment shown
in the setup diagram (CH 1/CH 2 slot). If the unit was in standby mode
set the ON/STANDBY switch to ON.
+%" +Initialize the mainframe as follows:
UTILITY button press...........................................
%* (in major menu) touch..............................
%* (in verification popĆup menu) touch..................
+%" +Set the horizontal record length as follows:
!#*! % $ (in major menu) touch.......................
!# % (in popĆup menu) touch.................
Set !# to 1024 points adjust upper knob.........
+%" +Identify the trace as follows:
# icon touch.............................................
Identify Channel () touch...................................
Select %# $ touch....................................
'" % &#)