Datasheet

Performance Verication
Frequency Amplitude
300 MHz
–27 dBm (~10 m V
RMS
)
2.5 GHz –21 dBm (~20 mV
RMS
)
3.0 GHz –15 dBm (~40 m
V
RMS
)
FCA3X20
250 MHz
–21 dBm (~20
mV
RMS
)
500 MHz
–21 dBm (~2
0mV
RMS
)
18 GHz –27 dBm (~
10 mV
RMS
)
20 GHz –21 dBm (~
20 mV
RMS
)
MCA30XX
300 MHz
–33 dBm
(~5 mV
RMS
)
18 GHz –33 dBm
(~5 mV
RMS
)
20 GHz –29 dB
m(~8mV
RMS
)
27 GHz –27 dB
m (~10 mV
RMS
)
40 GH
z
–23 d
Bm (~16 mV
RMS
)
Pus
htheUser Opt button, and then select Save Recall > Recall Setup > Default
to return the instrument to the factory default settings.
This completes the Performance Verication procedure