Performance Verification
Equipment Setu
p
Procedure
1. Connect the as shown. This is the same setup as the preceding step.
2. Power o n the instruments. Wait at least 20 minutes. (See page 2-8, Instrument
Warmup Require
ments.)
3. On the Tektronix FCA/MCA3000 Series instrument, push the Meas button
and select Freq > Freq > C.
4. Set the Microwave Signal Generator to output the first frequency listed below,
for the Tektronix FCA/MCA3000 Series instrument being tested.
5. Adjust the Microwave Signal Generator output so that the Power Meter shows
the amplitude listed for the frequency setting made in Step 4.
6. Enter Pass in the test record if the first four significant digits of the Tektronix
FCA/MCA3000 Series frequency readout match the microwave signal
generator’s frequency. Enter Fail if the frequency readouts do not agree.
Digits after the first 4 most significant do not matter.
7. Repeat steps 4 through 6 for the remaining frequencies shown for the
instrument being tested.
Frequency Amplitude
FCA3X03
100 MHz
–21 dBm (~20 mV
RMS
)
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 2–29