Datasheet

Performance Verication
Procedure
1. Connect the as s
hown:
FCA3100 Series instruments: Attach the type N to 2.92 mm adapter to
the FCA3100 Se
ries instrument. Do not connect the power splitter to it
for the initial s teps in the procedure. The power splitter is connected to
the adapter in Step 5.
MCA3000 Series instruments: Attach the 2.92 mm male to 2.92 mm
male adapter to the MCA3000 Series instrument. Do not connect the
power splitter to it for the initial steps in the procedure. The power splitter
is connect
ed to the adapter in Step 5.
2. Power o n the instruments. Wait at least 20 minutes. (See page 2-8, Instrument
Warmup Re
quirements.)
3. Set the Microwave signal source to output +7 dBm at 500 MHz.
4. Set the Power Meter to measure power at roughly +1 dBm relative to the
input that it currently is receiving. Use these steps to set a Rohde & Schwarz
NRVS Po
wer Meter:
a. Push W<->dBm twice, to display dBm.
b. Verify that the signal is approximately +1 dBm, and log the result in the
test record as Relative, under Input Impedance.
c. Push Unit.
d. Push REL, under the display.
e. Push ΔdB under the display.
f. Press STO under the display.
FCA3000, FCA3100, and MCA3000 Series Specications and Performance Verication 2–27