Datasheet
Performance Verification
8. Enter measured
frequency in the test report. (It should be close to
10.000010 MHz.)
9. Change the Function Generator frequency to 9.999990 MHz.
10. Enter measured frequency in the test report. (It should be close to
10.000010 MHz.)
11. Set the FCA/MCA3000 Series instrument to the factory defaults; push User
Opt > SaveRecall > Setup > Recall Setup > Default.
10 MHz Out Amplitude
Equipment:
Oscilloscope
BNC male to BNC male Cable (24 inches)
Equipment Setup
Procedure
1. Connect the as shown.
2. Power on the instruments. Wait at least 20 minutes.
3. Set the Oscilloscope:
a. Set channel 1 to 50 Ω input, 500 mV/div,andDC coupled.
b. Set the timebase to 40 nS/div.
c. Adjust the trigger for a stable display.
4. Verify that the signal amplitude on the oscilloscope is ≥1V
RMS
(≥2.828 V
P-P
).
Enter this amplitude in the test record.
5. Verify that the output frequency is approximately 10 MHz. At 40 nS/Div two
complete sine waves should have a duration of 5 divisions.
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 2–25