Datasheet
Performance Verification
External Refe
rence Clock Frequency and Sensitivity
Equipment:
Function Generator
BNC male to BNC male 50 Ω cable (24 inches) (Qty 2)
Equipment Setup
Procedure
1. Connect the equipment as shown.
2. Power on the instruments. Wait at least 20 minutes.
NOTE. To ensure that the external timebase is being used, the frequency is
intentionally offset. If the measured value does not agree with the expected value
shown in the procedure, the signal applied to the DUT External Reference input
may not be adequate.
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 2–23