Datasheet

Performance Verication
4. On the Tektroni
x FCA/MCA3X00 Series instrument:
a. Push the Meas button and select Time > Time Interval > A to B.
b. Push the Settings button, select Meas Time,enter20 ms, and push the
Save/Exit buttontoreturntothemainscreen.
c. Push the Input A button.
d. Set the input coupling to DC.
e. Set slope to Falling.
f. Set the input impedance to 50 .
g. Set the in
put attenuator to 1X.
h. Set the Trigger mode to Man.
i. Push the Trig menu button, enter 0.0 V, and push the Save/Exit button to
return to the main screen.
j. Push the Input B button.
k. Repeat steps d through i to set the Input B parameters.
l. Push the Value button.
5. On the Tektronix FCA/MCA3X00 Series instrument:
a. Push
the Hold/Run button to put the FCA/MCA3X00 Series instrument
into Hold mode.
b. Pus
htheValue button. This runs a single acquisition and returns to Hold
mode.
c. En
ter the time interval measurement value in the table below.
1. 6.
2.
7.
3. 8.
4. 9.
5.
1
0.
e. Repeat parts b and c nine times, for a total of 10 e ntries in the table above.
f. Subtract the smallest value from the largest, and enter the result in the test
record. U se the appropriate location for the instruments timebase option
(Std., Opt. MS, Opt. HS, or Opt. US).
2–22 FCA3000, FCA3100, and MCA3000 Series Specications and Performance Verication