Datasheet
Performance Verification
6. On the Function
Generator, at each of the following frequencies, verify that
the Tektronix FCA/MCA3X00 Series instrument is counting, and that the
count is accurate. Enter the frequency in the test record.
10 Hz
10 kHz
200 MHz
7. On the Function Generator, push the Output ON buttontoturnofftheoutput
signal (button goes from green to unlit).
8. On the Tektronix FCA/MCA3X00 Series instrument, disconnect the cable and
attenuator from Channel A and connect them to Channel B.
9. Repeat steps 4 through 7. Substitute Channel B for the instructions in steps
5aand5c.
10. Set the Microwave Signal Generator output to Off, set the output amplitude to
0dBm, and then set the output frequency to 300 MHz.
11. Disconnect the cable from the Function Generator and connect it to the
adapter on the Microwave Signal Generator.
12. Set the Microwave Signal Generator output to On.
13. Enter the Frequency shown on the Tektronix FCA/MCA3X00 Series
instrument as Channel B 300 MHz sensitivity in the test record.
14. Set the Microwave Signal Generator to 400 MHz.
15. Enter the Frequency shown on the Tektronix FCA/MCA3X00 Series
instrument as Channel B 400 MHz sensitivity in the test record.
16. On the Microwave Signal source, set the output to Off.
17. On the Tektronix FCA/MCA3X00 Series instrument:
a
.
D
isconnect the cable from Channel B and connect it to Channel A.
b. Push the Meas button and select Freq > Freq > A.
18. On the Microwave Signal source, set the output to On.
19. Enter the Frequency shown on the Tektronix FCA/MCA3X00 Series
instrument as Channel A 400 MHz sensitivity in the test record.
20. Set the Microwave Signal source to 300 MHz.
21. Enter the Frequency shown on the Tektronix FCA/MCA3X00 Series
instrument as Channel A 300 MHz sensitivity in the test record.
22. On the Microwave Signal source, set the output to Off.
2–18 FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification