Datasheet
Performance Verification
4. Set the Functio
n Generator to produce a 10 Hz sinewave at 150 mV
RMS
(424.2 mV
p-p
). Follow these steps to set a Tektronix AFG3251:
a. Push the Sine f
unction button.
b. Push the Frequency/Period menu button.
c. Push the Frequency menu button.
d. Use the keypad and menu buttons to set the output frequency to 10 Hz.
e. Push the Top Menu button.
f. Push the Amplitude/Level menu button.
g. Push the –
More- side menu button.
h. Push the Units side menu button, and then push the Vrms side menu
button.
i. Push the –More- side menu button.
j. Push the Amplitude menu button.
k. Use the keypad and menu buttons to set the output level to 150 mV
RMS
(424.2 mV
p-p
).
l. Push the Output ON button to turn on the signal (On button glows green).
5. On the Tektronix FCA/MCA3X00 Series instrument:
a. Push the Meas button and select Freq > Freq > A.
b. Pus
htheSettings button, select Meas Time,enter20 ms, and then push
the Save/Exit buttontoreturntothemainscreen.
c. Pu
sh the Input A button.
d. Set the input coupling to DC.
e. Set the input impedance to 50 Ω.
f. Set the input attenuator to 1X.
g. Set the Trigger mode to Man.
h. Push the Trig menu button, enter 0V, and then push the Save/Exit button
to return to the main screen.
i. Push the Analyze button as many times as needed, until you see the
Numerical Analysis display. (See Figure 2-1.)
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 2–17