9. Set the Functio
n Generator to generate a 100 Hz, 1 V
pp
signal. On an
AFG3251, use these steps:
a. Push the Sine f
unction button.
b. Push the Frequency/Period side menu button.
c. Push the Frequency side menu button.
d. Use the keypad and side m enu buttons to enter 100 Hz.
e. Push the Top Menu panel button.
f. Push the Amplitude/Level Menu side menu button.
g. Push Offs
et.
h. Use the keypad and side menu buttons to set the offset voltage to 0mV.
i. Push the Amplitude side menu button.
j. Use the keypad and side menu buttons to set the amplitude to 1V
pp
.
k. Use the arrow buttons below the control knob to position the control
cursor to the 10 mV unit of the amplitude value so that the knob changes
the output amplitude in 10 mV increments.
l. Push the Output ON buttontoturnonthesignaloutput.
10. Move the cable and attenuator stack from the feedthrough terminator on the
voltmeter to the Channel A input on the FCA/MCA3X00 Series instrument.
Both 10X attenuators are installed. (Do not include the 50 Ω feedthrough
termination.)
2–12 FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification