Datasheet
Performance Verification
Performance V
erification
This procedu
re tests the FCA/MCA3X00 Series instrument, to verify that it meets
all of the guaranteed specifications.
NOTE. Before starting the Performance Verification procedures, look at the
serial number sticker on the back of the FCA/MCA3X00 Series instrument to see
which timebase the instrument has; Std., Opt. MS, Opt. HS or Opt. US. Note
the timebase type on the test record.
Instrument Warmup
Requirements
All FCA/MCA3X00 Series models require a 48 hour warm-up period,except
FCA models with a Standard Timebase. FCA models with a Standard Timebase
only require a 20 minute warm-up.
Channel A and B Input Resistance
Equipment:
Digital Multimeter
Coaxial cable, BNC male to BNC male
BNCtoBananaPlugadapter
Equipment Setup
Procedure
1. Connect the equipment as shown.
2. Power on the instruments for at least 20 minutes.(See page 2-8, Instrument
Warmup Requirements.)
2–8 FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification