Datasheet

Performance Verication
Instrument perform ance test Minimum Measured Maximum
300 MHz @ 25 mV
RMS
297 MHz 303 MHz
400 MHz @ 25 mV
RMS
396 MHz 404 MHz
Timebase Ac
curacy
FCA3X00 Ser
ies
Std (7.00E
–6
)
0.9999930000 MHz 1.0000070000 MHz
Opt. MS (2.
40E
–7
)
0.9999997600 MHz 1.0000002400 MHz
Opt. HS (6
.00E
–8
)
0.9999999400 MHz 1.0000000600 MHz
MCA3X00
Series
Std (2.4
0E
–7
)
0.9999997600 MHz 1.0000002400 MHz
Opt. HS
(6.00E
–8
)
0.9999999400 MHz 1.0000000600 MHz
Opt. US
(1.80E
–8
)
0.9999999820 MHz 1.0000000180 MHz
Quant
ization Error
FCA3
100 Series
——
65 ps
RMS
FCA3
000 Series
——
100 ps
RMS
MCA
3000 Series
——
100 ps
RMS
Ex
ternal Reference Clock Frequency and Sensitivity
0.999999 MHz 1.000010 MHz
4.999995 MHz 1.000010 MHz
9
.999990 MHz
1
.000010 MHz
1
0 MHz O ut Amplitude
Output Amplitude
1.0 V
——
2–4 FCA3000, FCA3100, and MCA3000 Series Specications and Performance Verication