Datasheet
Specifications
Table 1-5: Guaranteed measurement capabilities (cont.)
Characteristics Description
1X Attenuator
DC, 1 Hz to 1 kHz: 1% +15 mV
1kHzto20MHz:
3% + 15 mV
20 MHz to 100 MHz:
10% + 15 mV
100 MHz to 300 MHz:
30% + 15 mV
10X Attenuator
DC, 1 Hz to 1 kHz: 1% +150 mV
1kHzto20MHz:
3% + 150 mV
20 MHz to 100 MHz:
10% + 150 mV
Accuracy (5 V r ange)
100 MHz to 300 MHz:
30% + 150 mV
Power
(MCA3000 Series channel C)
Product can measure the power input to channel C
Power R ange -35 dBm to +10 dBm
Frequency Range
MCA3027: 0.3 GHz to 27 GHz
MCA3040: 0.3 GHz to 40 GHz
Resolution 0.01 dB m at 100 ms measuring time
Accuracy
MCA3027: 0.3 GHz to 27 GHz: ±1 dBm
MCA3040: 0.3 GHz to 20 GHz: ±3 dBm
>27 GHz to 40 GHz: ±2 dBm
Totalize A, B (FCA31XX Only) The instrument is capable of totalizing counts
Totalize Modes Tot A, Tot B, Tot A+B, Tot A–B
Totalize Range 1 to 10
10
counts
Totalize Frequency Range < 160 MHz
Totalize Controls Start: Manual or Start Arming
Stop: Manual or Stop Arming or Timed
Ancilliary Measurements
Totalize A: A–B, A/B
Totalize B: A–B, A/B
Totalize A+B: A, B
Totalize A–B: A, B
Totalize A/B: A, B
Table 1-6: Other available readings
Characteristic
1
Description
Frequency Ratio
Able to provide the r atio of two input signals. Available measurements may depend upon
selected options
Range 10
-9
to 10
11
Lowest Input Frequency for
Correct Operation
0.1 Hz
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 1–13