Datasheet
Specifications
Table 1-5: Guaranteed measurement capabilities (cont.)
Characteristics Description
Positive and Negative
Pulse Wid th
Capable of measuring Pulse Width on A or B channels
Range 2.3 ns to 10
6
s
Minimum Pulse Width 2.3 ns
Total Measurement
Uncertainty
Where:
RMU = Random Measurement Uncertainty
SMU = Systemic Measurement Uncertainty
Random Measurement
Uncertainty
Where:
QE = Quantiz
ation Error
TE
X
= Trigger Error for a particular edge X
Systemic Measurement
Uncertainty
Where:
Etl = Trigger Level Timing Error
CTD = Channel Timing Difference
TBE = Timebase Error
MR = Measurement Result
Trigger Level Timing Error
(Etl)
Where:
TLU = Trigger Level Uncert
ainty
Hyst = Hysteresis W indow
S
X
= Slew R ate at the trigger point for timing location X
Ancillary Measurements V
max
,V
min
,V
p-p
Rise and Fall Time
Capable of measuring Rise or Fall Time on A or B channels
Range 1.5 ns to 10
6
sec.
Trigger Levels
10% and 90% of signal lev
el
Pulse Width 1.6 ns
Total Measurement
Uncertainty
Where:
RMU = Random Measurement Uncertainty
SMU = Systemic Measurement Uncertainty
1–10 FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification