Datasheet
Specifications
Table 1-5: Guaranteed measurement capabilities (cont.)
Characteristics Description
Random Measurement
Uncertainty
(FCA3000, FCA3100,
MCA3000, channels A
and B)
For measurement time ≥200 ms and Smart Freq = Auto or ON:
NOTE. This equation is the default o n FCA3100 Series.
Where:
MR = M easurement reading
QE = Quantization Error
TE = Trigger Error
MT = Measurement Time
N = 800/MT with the following limits: 6 ≤ N ≤ 1000 and N<((MR/2)*MT) – 2
For FCA3100, additionally, N < MT/8 μs
For measurement time <200 ms and Smart Freq = Auto or OFF:
Where:
MR = M easurement reading
QE = Quantization Error
TE = Trigger Error
MT = Measurement Time
Systemic Measurement
Uncertainty
(FCA3000, FCA3100,
MCA3000 channels A and B)
Where:
TBE = Timebase Error
MR = Measurement Reading
MT = Measurement Time
GTU = Gate Time Uncertainty
Ancillary Measurements Vmax, Vmin, V
p-p
Period
Capable of measuring period on any of up to three inputs
Mode
Single, Average
Range
Channel A, B: 3.3 ns to 1000 s (single, average)
Channel C (Option): 10 ns down to 330 ps, 125 ps, 70 ps, or 50 ps
Resolution
(FCA3000 Series)
100 ps (single shot) 12 digit/s (average)
FCA3000, FCA3100, and MCA3000 Series Specifications and Performance Verification 1–7