Datasheet

Auxiliary input Front-panel BNC connector. Input Impedance 1 MΩ. Max input 300 V
RMS
CAT II with peaks ≤ ±425 V.
Probe compensator output voltage
and frequency
Front-panel pins
Amplitude 0 to 2.5 V
Frequency 1 kHz
Auxiliary out Rear-panel BNC connector
V
OUT
(Hi): ≥2.5 V open circuit, ≥1.0 V 50 Ω to ground
V
OUT
(Lo): ≤0.7 V into a load of ≤4 mA; ≤0.25 V 50 Ω to ground
Output can be configured to provide a pulse out signal when the oscilloscope triggers, the internal oscilloscope reference clock
out, or an event out for limit/mask testing.
External reference input Time-base system can phase lock to an external 10 MHz reference (10 MHz ±1%)
Kensington-style lock Rear-panel security slot connects to standard Kensington-style lock.
VESA mount Standard (MIS-D 100) 100 mm VESA mounting points on rear of instrument.
LAN eXtensions for Instrumentation (LXI)
Class LXI Class C
Version V1.3
Power source
Power source voltage 100 to 240 V ±10%
Power source frequency 50 to 60 Hz ±10% at 100 to 240 V ±10%
400 Hz ±10% at 115 V ±13%
Power consumption 250 W maximum
Physical characteristics
Dimensions
mm
in.
Height 229 9.0
Width 439 17.3
Depth 147 5.8
Weight
kg
lb.
Net 5 11
Shipping 10.7 23.6
Rackmount configuration 5U
Cooling clearance 2 in. (51 mm) required on left side and rear of instrument
Datasheet
Input/output ports
20 www.tek.com