Datasheet
Digital Phosphor Oscilloscopes
DPO4000 Series
DPO4000 • www.tektronix.com/oscilloscopes
8
Characteristics
Vertical System
DPO4034 DPO4054 DPO4104
Input Channels 4 4 4
Analog Bandwidth 350 MHz 500 MHz 1 GHz
(–3 dB) 5 mV/div to 1 V/div)
Calculated Rise Time 5 mV/div (typical) 1 ns 700 ps 350 ps
Hardware Bandwidth Limits 20 MHz or 250 MHz
Input Coupling AC, DC, GND
Input Impedance 1 MΩ ±1%, 50 Ω ±1%
Input Sensitivity, 1 MΩ 1 mV/div to 10 V/div
Input Sensitivity, 50 Ω 1 mV/div to 1 V/div
Vertical Resolution 8 bits (11 bits with Hi Res)
Max Input Voltage, 1 MΩ 250 V
RMS
with peaks ≤±400 V
Max Input Voltage, 50 Ω 5V
RMS
with peaks ≤±20 V
DC Gain Accuracy ±1.5% with offset set to 0 V
Offset Range 1MΩ 50 Ω
1 mV/div to 50 mV/div ±1 V ±1 V
50.5 mV/div to 99.5 mV/div ±0.5 V ±0.5 V
100 mV/div to 500 mV/div ±10 V ±10 V
505 mV/div to 995 mV/div ±5 V ±5 V
1 V/div to 5 V/div ±100 V ±5 V
5.05 V/div to 10 V/div ±50 V NA
Channel-to-Channel Isolation ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
for any two channels having equal volts/div settings
Horizontal System
DPO4034 DPO4054 DPO4104
Maximum Sample Rate (all channels) 2.5 GS/s 2.5 GS/s 5 GS/s
Maximum Record Length (all channels) 10 M points
Maximum Duration at Highest 4 ms 4 ms 2 ms
Sample Rate (all channels)
Timebase Range (s/div) 1 ns to 400 s 400 ps to 400 s
Timebase Delay Time Range –10 divisions to 50 s
Channel-to-Channel Deskew Range ±100 ns
Timebase Accuracy ±5 ppm over any ≥1 ms interval