User's Manual
Commands Listed in Alphabetical Order
DESkew:DISPl
ay
Sets or returns the state of the deskew table display.
Group
Vertical
Syntax
DESkew:DISPlay {OFF|ON|0|1}
DESkew:DISPlay?
Arguments
OFF or 0 turns off the deskew table d isplay.
ON or 1 turns on the deskew table display.
DIAg:LOOP:OPTion
Sets the self-test loop option.
Group
Calibration a nd Diagnostic
Syntax
DIAg:LOOP:OPTion {ALWAYS|FAIL|ONFAI L|ONCE|NTIMES}
Arguments
ALWAYS continues looping until the self tests (diagnostics) are stopped via the
front panel or by an oscilloscope command.
FAIL causes looping until the ļ¬rst self test (diagnostic) failure or until self tests
(diagnostics) are stopped.
ONFAIL causes looping on a spe ciļ¬ c test group as long as a FAIL status is
returned from the test.
ONCE executes se lf test (diagnostics test) sequence once.
NTIMES runs ānā number of loops.
Examples
DIAG:LOOP:OPTION ONCE runs one loop of self tests.
DIAg:LOOP:OPTion:NTIMes
Sets the self-test loop option to run N times.
Group
Calibration a nd Diagnostic