Datasheet
Data Sheet
White edges indicate additional information is available by zooming in. As shown here,
zooming in on the white edge reveals a hidden glitch.
The MagniVu high-resolution record provides 121.2 ps timing resolution, enabling you to
make critical timing measurements on your digital waveforms.
The multiple transition detection hardware of the MSO3000 Series will
show you a white edge on the display whe n the system detects multiple
transitions. White edges indicate that more information is available by
zooming in or acquiring at faster sampling rates. In most cases zooming in
will reveal the pulse that was not viewable with the previous settings. If the
white edge is still present after zoomi
ng in as far as possible, this indicates
that increasing the sample rate on the next a cquisition will reveal higher
frequency information than the previous settings could acquire.
The MSO3000 Series simplifies channel setup by allowing you to group
digital waveforms and enter waveform labels by using a USB keyboa rd.
By simp
ly placing digital waveforms next to each other, they form a group.
The P6316 MSO probe offers two eight-channel pods to simplify connecting to your
device.
Once a group is formed, you can position all the channels contained in that
group collectively. This greatly reduces the normal setup time associated
with positioning channels individually.
MagniVu™ High-speed Acquisition
The main digital acquisition mode on the MSO3000 Series will capture up
to 5 Mpoints at 5 00 MS/s (2 ns resolution). In addition to the main record,
the MSO30
00 provides an ultra high-resolution record called MagniVu which
acquires 10,000 points at up to 8.25 GS/s (121.2 ps resolution). Both main
and MagniVu waveforms are acquired on every trigger and can be switched
between in the display at any time, running or stopped. MagniVu provides
significantly finer timing resolution th an comparable MSOs on the market,
instilling confidence when making critical timing measurements on digital
wavefor
ms.
P6316 MS
O Probe
This unique probe design offers two eigh t-chann el pods, simplifying the
process
of connecting to the device-under-test. Whe n conn ecting to square
pins, the P6316 can connect direc tly to 8×2 square pin headers spaced
on tenth-inch centers. When m ore attachment flexibility is required, you
can use the included flying lead set s and grabbers to clip onto surface
mount devices or test points. The P6316 offers ou tstanding electrical
characteristics applying only 8 pF of capacitive loading with 101 k input
impeda
nce.
6 www.tektronix.com