Datasheet
Digital Phosphor Oscilloscopes — MSO3000 Series, DPO3000 Series
Characteristics
Vertical Syst
em Analog Channels
Characteristic MSO3012
DPO3012
MSO3014
DPO3014
MSO3032
DPO3032
MSO3034
DPO3034
MSO3054
DPO3054
Input Channels
24244
Analog Bandwidth (-3 dB)
100 MHz 100 MHz 300 MHz 300 MHz 500 MHz
Calculated Rise Time
5 mV/div (typical)
3.5 ns 3.5 ns 1.17 ns 1.17 ns 700 ps
Hardware Ba
ndwidth Limits
20 MHz 20 MHz, 150 M
Hz
Input Coupling AC, DC, GND
Input I mpe
dance
1M ±1%, 75 ±1%, 50 ±1%
Input Sensitivity Range, 1 M 1mV/divto10V/div
Input Sensitivity Range, 75 ,50 1mV/divto1V/div
Vertical R
esolution
8 bits (11 bits with Hi Res)
Maximum Input Voltage, 1 M
300 V
RMS
w
ith peaks ±450 V
Maximum Input Voltage, 75 ,50
5V
RMS
wit
h peaks ±20 V
DC Gain Accuracy ±1.5% for 5 m V /div and above
±2.0% for 2 mV/div
±2.5% for 1 mV/div
Channel-to-Channel Isolation
(Any Two Channels at Equal Vertical Scale)
100:1 at 100 MHz and 30:1 at >100 MHz up to the rated BW
Offset Range
Range
1MΩ 50 Ω,75Ω
1mV/d
iv to 99.5 mV/div
±1 V ±1 V
100 mV
/div to 995 mV/div
±10 V ±5 V
1V/di
v
±100 V ±5 V
1.01 V
/div to 10 V/div
±100 V NA
Vertical System Digital Channels
Characteristic All MSO 3000 Mode ls
Input Channels 16 Digital (D15 to D0)
Thresholds
Threshold per set of 8 channels
Threshold Selections TTL, CMOS, ECL, PECL, User Defined
User-defined Threshold
Range
-15 V to +25 V
Maximum Input Voltage -20 V to +30 V
Threshold Accuracy
±(100 mV +3% of threshold setting)
Maximum Input Dynamic
Range
50 V
pk-pk
(threshold setting dependent)
Minimum Voltage Swing
500 mV
pk-pk
Input Impedance
101 k
Probe Loading 8 pF
Vertical Resolution 1 bit
Horizontal System Analog Channels
Characteristic All M SO 3000 Models
All DPO3000 Models
Maximum Sample Rate
(all channels)
2.5 GS/s
Maximum Record Length
(all channels)
5 Mpoints
Maximum Duration of
Time Captured at Highest
Sample Rate
(all channels)
2ms
Time-base Range (s/div)
1nsto1000s
Time-base Delay Time
Range
-10 divisions to 5000 s
Channel-to-Channel
Deskew Range
±100 ns
Time-base Accuracy ±10 ppm over any 1 ms interval
Horizontal System Digital Channels
Characteristic All M SO 3000 Models
Maximum Sample Rate
(Main, all channels)
500 MS/s (2 ns resolution)
Maximum Record Length
(Main, all channels)
5 Mpoints
Maximum Sample Rate
(MagniVu, all channels)
8.25 GS/s (121.2 ps resolution)
Maximum Record Length
(MagniVu, all channels)
10 kpoints centered on the trigger
Minimum Detectable
Pulse Width
2.0 ns
Channel-to-Channel
Skew
500 ps typical
www.tektronix.com 11