Datasheet

Vertical system digital channels
Thresholds Threshold per set of 8 channels
Threshold selections TTL, CMOS, ECL, PECL, User-defined
User-defined threshold range –15 V to +25 V
Maximum input voltage –20 V to +30 V
Threshold accuracy ±[100 mV + 3% of threshold setting]
Maximum input dynamic range 50 V
p-p
(threshold setting dependent)
Minimum voltage swing 500 mV
p-p
Input resistance 101 kΩ
Probe loading 8 pF
Vertical resolution 1 bit
Horizontal system analog channels
Maximum duration of time
captured at highest sample rate
(all channels)
2 ms
Seconds/division range 1 ns/div to 1000 s/div
Time-base delay time range -10 divisions to 5000 s
Channel-to-channel deskew range ±100 ns
Time base accuracy ±10 ppm over any ≥1 ms interval
Horizontal system digital channels
Seconds/division range 1 ns/div to 1000 s/div
Maximum record length (main, all
channels)
5 M points
Maximum sample rate (MagniVu,
all channels)
8.25 GS/s (121.2 ps resolution)
Maximum record length (MagniVu,
all channels)
10 k points centered on the trigger
Minimum detectible pulse width 2.0 ns
Channel-to-channel skew 500 ps, typical, digital channel to digital channel only This is the propagation path skew and ignores skew contributions due to
bandpass distortion, threshold inaccuracies (see Threshold accuracy), and sample binning (see Digital channel timing resolution)
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes