Datasheet
TEK.COM 9
DMM7512 Dual Channel 7½-Digit Sampling Multimeter
RMS Noise (additional peak noise uncertainty)
33
Applies to ± ppm of range.
Peak noise uncertainty is included in DC specifications for ≥1 PLC.
Add peak noise uncertainty to measurements when <1 PLC.
Examples 1 mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range.
1 mA at 1 PLC: 9 +0 = 9 ppm of range.
NPLC Digits 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A
5 7½ 0.15 0.14 0.09 0.1 0.3 0.3 0.2
1 7½ 0.4 0.13 0.1 0.1 0.5 0.5 0.3
0.2 6½ 0 (220) 0 (23) 0.2 (3.4) 0.2 (1.6) 2 (10) 2 (11) 0.7 (4.6)
0.2
34
6½ 120 (260) 12 (26) 1.2 (3.8) 0.3 (1.8) 1.9 (9.8) 2 (10) 0.8 (5)
0.06 5½ 130 (280) 12 (29) 1.3 (5.6) 0.4 (3.9) 2 (14) 2 (14) 1.2 (7.7)
0.006 4½ 130 (350) 14 (42) 3 (20) 2 (20) 4 (30) 4 (31) 7 (51)
0.0005 3½ 26 0 (2110) 30 (300) 20 (150) 20 (160) 30 (190) 30 (190) 70 (510)
DC Current Characteristics
Range 10 µA 100 µA 1 mA 10 mA 100 mA 1 A 3 A
Effective Internal Shunt Value
35
1 kΩ 100 Ω 10 Ω 1 Ω 0.1 Ω 0.1 Ω 0.1 Ω
Autozero Off Error: For ±1 °C and ≤10 minutes
add ±(8 ppm of reading + range error)
150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA
Overload Recovery: For each additional sustained amp
beyond ±1.5 A, add the following initial ppm of range error
until thermally settled after overload recovery
15500 1800 150 150 6500 200 —
Temperature
4-Wire RTD or 3-Wire RTD
Types: 100 Ω platinum PT100, D100, F100, PT385, PT3916; or user-configurable 0 Ω to 10 kΩ.
Type Range Resolution
Accuracy ± °C
2 Year, T
CAL
±5 °C Temperature Coefficient
36
4-Wire RTD –200° to 400 °C 0.01 °C 0.09 °C 0.003 °C/ °C
3-Wire RTD
37
–200° to 400 °C 0.01 °C 0.75 °C 0.003 °C/ °C
Thermistor
Types: 2.252 kΩ, 5 kΩ, and 10 kΩ.
Type Range Resolution
Accuracy ± °C
2 Year, T
CAL
±5 °C Temperature Coefficient
36
Thermistor –80° to +150 °C 0.01 °C 0.08 °C 0.002 °C/ °C
NOTES
33. Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is guaranteed.
34. With line sync on.
35. Values are typical and guaranteed by design.
36. Add per degree from T
CAL
±5 °C; specifications without autocalibration.
37. For 3-wire RTD, accuracy is for < 0.1 Ω lead resistance mismatch for input HI and LO. Add 0.25 °C/ 0.1 Ω of HI-LO lead resistance mismatch.