Datasheet

Datasheet
TEK.COM8
Resistance Characteristics
Maximum 4-Wire Ohms Lead Resistance
5 Ω per lead for 1 Ω range, 10% of range per lead for 10 Ω to 1 kΩ ranges; 1 kΩ per lead for all
other ranges.
Offset Compensation Selectable on 4-wire, 1 Ω to 100 kΩ ranges.
Open Lead Detector Default is off.
Autozero Off Error For 2-wire ohms, ±1 °C and 10 minutes, add ±(8 ppm of reading) + 1.5 mΩ for 10 Ω range, 15 mΩ for
100 Ω and 1 kΩ ranges, 150 mΩ for 10 kΩ range, 1.5 Ω for 100 kΩ range, and 15 Ω for all other ranges.
For 4-wire ohms, ±1 °C and 10 minutes, add ±(8 ppm of reading).
Input Current Limit For signals with a magnitude of +12 V to +40 V or –12 V to –40 V: ±13 mA source or sink, typical.
For signals with a magnitude of greater than +40 V or –40 V: ±130 µA source or sink, typical.
DC Current
Enhanced Accuracy (within 30 days of autocalibration, T
OPER
±5 °C from T
ACAL
)
Range
30
Resolution
Maximum Burden
Voltage
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1 °C
31
90 Day
T
CAL
±5 °C
1 Year
T
CAL
±5 °C
2 Year
T
CAL
±5 °C
Temperature
Coefficient
32
10.000000 µA 1 pA 15 mV 30 + 30 75 + 30 75 + 30 75 + 30 0.15 + 0.1
100.00000 µA 10 pA 15 mV 20 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0 .1
1.0000000 mA 100 pA 15 mV 30 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
10.000000 mA 1 nA 20 mV 40 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
100.00000 mA 10 nA 200 mV 50 + 18 150 + 30 150 + 30 150 + 30 0.15 + 0.1
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 400 + 50 400 + 50 0.15 + 0.1
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 400 + 40 400 + 40 0.15 + 0.1
Accuracy
Range
30
Resolution
Maximum Burden
Voltage
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1 °C
31
90 Day
T
CAL
±5 °C
1 Year
T
CAL
±5 °C
2 Year
T
CAL
±5 °C
Temperature
Coefficient
32
10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8
100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3
1.0000000 mA 100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3
10.000000 mA 1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3
100.00000 mA 10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10
NOTES
30. 20% overrange supported for all ranges except for 3 A, which is 1% supported.
31. Relative to calibration accuracy.
32. Add per degree from T
CAL
±5 °C.