Datasheet
Datasheet
TEK.COM6
Resistance
Enhanced Accuracy (within 30 days of autocalibration, T
OPER
± 5 °C from T
ACAL
)
9
Range
10
Resolution
Test Cur rent
11
(±5%)
Accuracy ±[ppm of reading + ppm of range]
24 Hour
T
CAL
±1 °C
12
90 Day
T
CAL
±5 °C
1 Year
T
CAL
±5 °C
2 Year
T
CAL
±5 °C
Temperature
Coefficient
13
1.0000000 Ω 0.1 µΩ 10 mA 15 + 60 30 + 60 30 + 60 30 + 60 0.15 + 0 .1
10.000000 Ω 1 µΩ 10 mA 15 + 6 30 + 6 30 + 6 30 + 6 0.15 + 0 .1
100.00000 Ω 10 µΩ 1 mA 12 + 4 27 + 4 27 + 4 27 + 4 0.15 + 0.1
1.0000000 kΩ 100 µΩ 1 mA 12 + 3 24 + 3 24 + 3 24 + 3 0.15 + 0.1
10.000000 kΩ
14
1 mΩ 100 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.15 + 0.1
100.00000 kΩ
14, 15
10 mΩ 10 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.3 + 0.1
1.0000000 MΩ
14, 16
100 mΩ 10 µA 14 + 3 30 + 4 30 + 4 30 + 4 0.7 + 0.1
10.000000 MΩ
17
1 Ω 0.69 µA || 10 MΩ 150 + 6 200 + 10 200 + 10 200 + 10 70 + 1
100.00000 MΩ
17
10 Ω 0.69 µA || 10 MΩ 800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1
1.0000000 GΩ
17
100 Ω 0.69 µA || 10 MΩ 9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1
Accuracy
18
Range
19
Resolution
Test Cur rent
20
(±5%)
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1 °C
21
90 Day
T
CAL
±5 °C
1 Year
T
CAL
±5 °C
2 Year
T
CAL
±5 °C
Temperature
Coefficient
22
1 Ω 0.1 µ Ω 10 mA 15 + 60 40 + 60 50 + 60 70 + 60 2.5 + 5
10 Ω 1 µΩ 10 mA 15 + 6 40 + 6 50 + 6 70 + 6 2.5 + 0.5
100 Ω 10 µΩ 1 mA 12 + 4 35 + 4 47 + 4 65 + 4 5 + 0.25
1 kΩ 100 µΩ 1 mA 12 + 3 30 + 3 41 + 3 65 + 3 5 + 0.25
10 kΩ
23
1 mΩ 100 µA 10 + 3 30 + 3 42 + 3 65 + 3 2.5 + 0.25
100 kΩ
23, 24
10 mΩ 10 µA 13 + 3 38 + 3 50 + 3 65 + 3 5 + 1
1 MΩ
23, 25
100 mΩ 10 µA 14 + 3 38 + 5 50 + 5 65 + 5 5 + 1
10 MΩ
26
1 Ω 0.69 µA || 10 MΩ 150 + 6 200 + 10 400 + 10 600 + 12 70 + 1
100 MΩ
26
10 Ω 0.69 µA || 10 MΩ 800 + 30 2000 + 30 2000 + 30 2600 + 30 385 + 1
1 GΩ
26
100 Ω 0.69 µA || 10 MΩ 9000 + 200 9000 + 200 13000 + 200 14000 + 200 3000 + 1
NOTES
9. Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ measurements. 1 Ω range is 4-wire only. For 2-wire,
with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add 100 mΩ to ppm of range uncertainty.
10. 20% overrange on all ranges.
11. Test current with offset compensation off.
12. Relative to calibration accuracy.
13. Add per degree from T
CAL
±5 °C.
14. Specifications are for external cable and load capacitance <1 nF.
15. For offset compensation on, add 10 ppm uncertainty to ppm of reading.
16. For 4-wire 1 MΩ, open lead detector on, add 10 ppm uncertainty to ppm of reading.
17. Specified for <10% lead resistance mismatch in HI and LO.
18. Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ measurements. 1 Ω range is 4-wire only. For 2-wire,
with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add 100 mΩ to ppm of range uncertainty.
19. 20% overrange on all ranges.
20. Test current with offset compensation off.
21. Relative to calibration accuracy.
22. Add per degree from T
CAL
±5 °C.
23. Specifications are for external cable and load capacitance < 1 nF.
24. For offset compensation on, add 10 ppm of uncertainty to ppm of reading.
25. For 4-wire, 1 MΩ, open lead detection on, add 10 ppm uncertainty to ppm of reading.
26. Specified for <10% lead resistance mismatch in HI and LO.