Datasheet

TEK.COM 15
DMM7512 Dual Channel 7½-Digit Sampling Multimeter
Analog Triggering
62
Analog Level, Edge, or Window Trigger Types
63
Trigger Characteristics Voltage Input Current Input
Input Range 100 mV to 1000 V 10 µA to 3 A
Resolution 0.05% 0.05%
Basic Accuracy (T
ACAL
±5 °C)
64, 65
1% 1%
Analog Trigger Latencies
Digital I/O External
Positive Logic 800 ns + 40 ns jitter 930 ns + 40 ns jitter
Negative Logic 800 ns + 40 ns jitter 840 ns + 40 ns jitter
Window Filter and Memory (buffer)
Window Filter Size 0 to 10% of reading, where 0 averages all readings.
Memory Up to 27.5 million timestamped readings with the compact buffer style, with additional memory available
using an external USB flash drive.
Maximum Internal Memory (Buffer) 27.5 million readings with the compact buffer style (6½-digit without formatting), 11 million readings with
the standard or full buffer style.
NOTES
62. For DC coupled, the trigger level can be set up to 100% of measure range.
63. Rising or falling edge triggering supported. Window trigger requires setting two independent levels.
64. Trigger event occurs after the threshold crossing at a time determined by total trigger latencies.
65. Accuracy specifications require user ACAL and are verified with level trigger amplitude set to 50% of range with a 100 Hz sine wave at 100% full scale of range. High frequency rejection is
off. NPLC 0.0005 (DC voltage/DC current) or aperture 1 µs for digitize voltage or digitize current. Specified for fixed range, autozero off. For DC current and digitized DC current 3 A range,