Datasheet

Digital Multimeters Tektronix DMM4050 and DMM4040
Electrical Specications
DC Voltage Specications
Accuracy specications are valid for digit resolution mode after at least a 1 hour
warm-up with Auto Zero enabled.
24-hour specications are relative to calibration standards and assume a controlled
electromagnetic environment per EN 61326-1:2000-11.
Characteristic
Description
Maximum Input 1000 V on any range
Common Mode
Rejection
140 dB at 50 or 60 Hz ±0.1% (1 k unbalance)
Normal Mode Rejection
60 dB for NPLC of 1 or greater with analog lter off and
power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog lter on and
power line frequency ±0.1%
Measurement Method
Multiramp A/D
A/D Linearity 0.0002% of measurement + 0.0001% of range
Input Bias Current <30 pA at 25 °C
Auto-zero Off Operation
Following instrument warm-up at calibration temperature
±1 °C and less than 10 minutes, add error: 0.0002%
range additional error +5 V
Analog Filter
When using the analog lter, specications are relative
towithinonehourofusingtheZEROfunctionforthat
range and NPLC setting
DC Ratio Accuracy is ±(Input accuracy + Reference accur acy),
where Input accuracy = DC Voltage accuracy for the HI
to LO Input (in ppm of the Input voltage), and Reference
accuracy = DC Voltage accuracy for the HI to LO
(Sense) Reference (in ppm of the Reference voltage)
Settling Considerations Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
Input Characteristics
ResolutionRange Resolution
Digits Digits Digits
Input Impedance
100 mV 100.0000 mV 10 V1V 100 nV
10 M or >10 G*
2
1 V 1.000000 V 100 V10V1V
10 M or >10 G*
2
10 V 10.00000 V 1 mV 100 V10V
10 M or >10 G*
2
100 V 100.0000 V 10 mV 1 mV 100 V
10 M ±1%
1000 V 1,000.000 V 100 mV 10 mV 1 mV
10 M ±1%
*
2
Inputs beyond ±14 V are c lamped through 200 k typical. 10 M is default input impedance.
www.tektronix.com/dmm405 0_4040 5