Datasheet
Data Sheet
www.tektronix.com/dmm4050_40406
DMM4050 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
100 mV 0.0025 + 0.003 0.0025 + 0.0035 0.0037 + 0.0035 0.0005 + 0.0005
1 V 0.0018 + 0.0006 0.0018 + 0.0007 0.0025 + 0.0007 0.0005 + 0.0001
10 V 0.0013 + 0.0004 0.0018 + 0.0005 0.0024 + 0.0005 0.0005 + 0.0001
100 V 0.0018 + 0.0006 0.0027 + 0.0006 0.0038 + 0.0006 0.0005 + 0.0001
1000 V 0.0018 + 0.0006 0.0031 + 0.001 0.0041 + 0.001 0.0005 + 0.0001
DMM4040 Accuracy
Accuracy is given as ±(% measurement + % of range)
Range 24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
100 mV 0.003 + 0.003 0.004 + 0.0035 0.005 + 0.0035 0.0005 + 0.0005
1 V 0.002 + 0.0006 0.003 + 0.0007 0.004 + 0.0007 0.0005 + 0.0001
10 V 0.0015 + 0.0004 0.002 + 0.0005 0.0035 + 0.0005 0.0005 + 0.0001
100 V 0.002 + 0.0006 0.0035 + 0.0006 0.0045 + 0.0006 0.0005 + 0.0001
1000 V 0.002 + 0.0006 0.0035 + 0.0010 0.0045 + 0.0010 0.0005 + 0.0001
Additional Errors
Digits NPLC Additional NPLC Noise Error
6½ 100 0% of range
6½ 10 0% of range
5½ 1 0.001% of range
5½ 0.2 0.0025% of range ±12 μV
4½ 0.02 0.017% of range ±17 μV