Data Sheet
Capacitance (DMM4050 Only)
Accuracy is stated as ±(% of measurement + % of range)
Range Resolution
1 Year Accuracy*
12
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside18to28°C
1nF 1pF
2% ±2.5%
0.05 + 0.05
10 nF 10 pF
1% ±0.5%
0.05 + 0.01
100 nF 100 pF
1% ±0.5%
0.01 + 0.01
1 F1nF
1% ±0.5%
0.01 + 0.01
10 F10nF
1% ±0.5%
0.01 + 0.01
100 F100nF
1% ±0.5%
0.01 + 0.01
1mF 1F
1% ±0.5%
0.01 + 0.01
10 mF 10 F
1% ±0.5%
0.01 + 0.01
100 mF 100 F
4% ±0.2%
0.05 + 0.05
*
12
Stated ac
curacy is attained when Zero function is used.
Te mp er ature (DMM4050 only)
Test Cu rrent: 1 mA
Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 lead resistance. The accuracy listed in the table below are valid
only when using the 4-wire RTD measurement function. Specifications do not include probe accuracy, which must be added.
AccuracyRange Resolutio
n
90 Days
(23°C±5°C
)
1Year
(23 °C ±5 °C
)
Temperature Coefficient/°C
Outside18to28°C
-200 °C 0.001 °C
0.06 0.09 0.0025
-100 °C 0.001 °C
0.05 0.08 0.002
0 °C 0.001 °C
0.04 0.06 0.002
100 °C 0.001 °C
0.05 0.08 0.002
300 °C 0.001 °C
0.1 0.12 0.002
600 °C 0.001 °C
0.18 0.22 0.002
Additional Errors
Digits
NPLC Additio
nal NPLC Noise Error
6½
100
0°C
6½
10
0°C
5½
1
0.03 °C
5½
0.2
0.12 °C
4½
0.02
0.6 °C
14 www.tektronix.com/dmm4050_4040