Datasheet
Data Sheet
www.tektronix.com6
AC Voltage Specifications
AC Voltage specifications are for AC sinewave signals >5% of range. For inputs
from 1% to 5% of range and <50 kHz, add an additional error of 0.1% of range,
and for 50 kHz to 100 kHz, add 0.13% of range.
Characteristic Description
Maximum Input 750 V
RMS
or 1000 V peak or 8 × 10
7
Volts-Hertz product
Measurement Method AC-coupled true RMS. Measures the AC component of
input with up to 1000 VDC bias on any range
AC Filter Bandwidth 20 Hz – 100 kHz
Common Mode Rejection 60 dB at 50 Hz or 60 Hz (1 kΩ unbalance)
Maximum Crest Factor 3:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1–2, 0.05% of full scale
Crest Factor 2–3, 0.2% of full scale
Only applies for non-sinusoid signals
Input Characteristics
Range Full Scale
(5½ Digits)
Resolution Input Impedance
Slow Medium Fast
200 mV 199.999 mV 1 μV 10 μV 10 μV 1 MΩ ±2% shunted by
<100pf
2 V 1.99999 V 10 μV 100 μV 100 μV
20 V 19.9999 V 100 μV 1000 μV 1000 μV
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
Accuracy
Range Frequency Uncertainty
*3
Temperature Coefficient/°C
Outside 18– 28°C
90 days 1 year
23°C ±5°C 23°C ±5°C
200 mV 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 +0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
2 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 +0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
20 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
200 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
750 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
*3
Uncertainty given as ±(% of reading + % of range).